Medical Product Specialist Inc dba MPS Acacia: Acacia Catalog # 385165 "Y" IV Extension Set 20cm Macro Bore, Spin Nut w/BD Q-Syte
- Recall number
- Z-1079-2010
- Recalling firm
- Medical Product Specialist Inc dba MPS Acacia
- Product
- Acacia Catalog # 385165 "Y" IV Extension Set 20cm Macro Bore, Spin Nut w/BD Q-Syte
- Status
- Terminated
- Initiated
- 2009-12-18
- Posted
- 2010-03-30
- Terminated
- 2011-01-05
- Reason
- Some lots of IV Extension Sets with BD Q-Syte Luer Access Split Septum have a manufacturing deviation which could cause the product to not function properly, possibly resulting in an air embolism.
- Root cause
- Process design
- Action
- A letter dated December 18, 2009 was sent to Acacia customers with the following instructions. 1. Immediately review your inventory of Acacia IV Extension Sets with BD Q-Syte for REF numbers and lot numbers listed and remove them from all inventory printed on the unit packaging , shelf box and shelf shipper containers. 2. Complete the Distributor Recall response Card (Attachment 1) and return by fax immediately to the number on the Distributor Recall Response Card for product in your facility at the time of receipt of this letter. This form must be completed and f axed to Acacia, Inc even if there is no product in inventory. 3. Acacia, Inc is currently working to ensure that replacement products are readily available. Once you have identified how much product you will be returning, please call your Customer Representative at 1-800-453-4538, opt. 2, ext. 2534, to arrange for product replacement. Questions are to be directed to the Customer Representative at 1-800-453-4538, opt. 2, ext. 2534, between 8:00AM and 5:00PM MST, Monday-Friday.
- Quantity
- 69,726
- Distribution
- US Only
- Lot, serial or model codes
- Lots A2100 and A2101
- 510(k) numbers
- ["K895367"]
- PMA numbers
- not on file
- Product code
- FPA
- Firm FEI
- 1419937
- Firm city
- Brea
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28