Sekisui Diagnostics P.E.I. Inc.: Sekisui Diagnostics Ammonia L3K Assay; Catalog Number: 293-80-91 and 293-10 For the IN VITRO quantitative measurement of ammonia concentration in plasma

Recall number
Z-1078-2017
Recalling firm
Sekisui Diagnostics P.E.I. Inc.
Product
Sekisui Diagnostics Ammonia L3K Assay; Catalog Number: 293-80-91 and 293-10 For the IN VITRO quantitative measurement of ammonia concentration in plasma
Status
Terminated
Initiated
2016-11-29
Posted
2017-01-23
Terminated
2017-07-06
Reason
Certain lots are showing an atypical decrease in optical density (OD) over time, which may result in an impact to performance at the high end of the linear range.
Root cause
Material/Component Contamination
Action
Sekisui Diagnostics sent an Important Product Correction Notification letter dated November 29, 2016, to all affected customers. Customers were asked to take the following actions: Customer Instructions: Please stop using and discard any of the Ammonia L3K reagent lots listed above in your inventory; Contact your distributor to discuss replacement product. Distributor Instructions: Please send the "IMPORTANT PRODUCT CORRECTION Customer Notification" provided to all customers that received Sekisui Diagnostics Ammonia LJK from the listed lot numbers; Complete, sign and return the attached CONFIRMATION OF NOTIFICATION form by FAX to the number provided. This will: Indicate receipt of this Field Correction, and confirmation of notification of customers that received impacted customers; Act as confirmation of destruction of affected product in inventory; Allow for issuance of replacement product. Customers with questions were instructed to contact Technical Services at 1-800-565-0265 or email PEIDiagnosticTechnical@sekisui-dx.com.
Quantity
774 Liters bulk and 565 kits
Distribution
Worldwide Distribution - US including DE, KS, FL, TX, CA, MI, TN, MA, ID, NJ, RI, IL and Internationally to Austria, Mexico, Canada, India, South Korea, China, Spain, South Korea, and Sri Lanka
Lot, serial or model codes
Catalog Number: 293-80-91; Lot Numbers: 48737, 48962, 49087, 49220, 49258 and Catalog Numbers: 293-10; Lot Numbers: 48967, 49330, 49331, 49515
510(k) numbers
["K033921"]
PMA numbers
not on file
Product code
JIF
Firm FEI
3002806809
Firm city
Charlottetown
Firm state
not on file
Firm country
Canada

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register