Skip to the content

FDA Medical Device Recalls (openFDA)

Dental Equipment LLC: DCI Equipment Track Mount Light Assembly, 110v

Recall number
Z-1077-2007
Recalling firm
Dental Equipment LLC
Product
DCI Equipment Track Mount Light Assembly, 110v
Status
Terminated
Initiated
2007-05-01
Posted
2007-07-12
Terminated
2007-07-12
Reason
Potential for light shield to melt due to light assemblies not properly rated to handle the VAC.
Root cause
Other
Action
On 5/1/07 the dental office contacted the manufacturer via telephone regarding the incident. On 5/2/07 the firm's service representative visited the dental office and replaced the lights.
Quantity
4 dual track lights
Distribution
The lights were distributed to one distributor in NY who in turn distributed the lights to one dental office.
Lot, serial or model codes
All products of this type. Product does not contain serial numbers or any other identifying codes.
510(k) numbers
not on file
PMA numbers
not on file
Product code
EAZ
Firm FEI
1000523896
Firm city
Newberg
Firm state
OR
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register