Skip to the content

FDA Medical Device Recalls (openFDA)

INO Therapeutics (dba Ikaria): Mallinckrodt Pharmaceuticals INOmax DSIR Plus delivery system. Model 10007. Delivers INOMAX (nitric oxide for inhalation). The INOmax DSIR uses a dual-channel design to ensure the safe delivery of INOMAX. The first channel has the delivery CPU, the flow controller and the injector module to ensure the accurate delivery of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO2, and 02 cells) and the user interface including the display and alarms. The dual-channel approach to delivery and monitoring permits INOMAX delivery independent of monitoring but also allows the monitoring system to shutdown INOMAX delivery if it detects a fault in the delivery system such that the NO concentration could become greater than 100 ppm. The delivery system can also shut down delivery if it detects certain serious problems with the monitoring system.

Recall number
Z-1076-2017
Recalling firm
INO Therapeutics (dba Ikaria)
Product
Mallinckrodt Pharmaceuticals INOmax DSIR Plus delivery system. Model 10007. Delivers INOMAX (nitric oxide for inhalation). The INOmax DSIR uses a dual-channel design to ensure the safe delivery of INOMAX. The first channel has the delivery CPU, the flow controller and the injector module to ensure the accurate delivery of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO2, and 02 cells) and the user interface including the display and alarms. The dual-channel approach to delivery and monitoring permits INOMAX delivery independent of monitoring but also allows the monitoring system to shutdown INOMAX delivery if it detects a fault in the delivery system such that the NO concentration could become greater than 100 ppm. The delivery system can also shut down delivery if it detects certain serious problems with the monitoring system.
Status
Terminated
Initiated
2016-11-21
Posted
not on file
Terminated
2017-12-20
Reason
Mallinckrodt discovered that some devices may have been manufactured with an out-of-tolerance manometer which can result in in an inaccurate proportional valve calibration table in an affected device.
Root cause
Equipment maintenance
Action
Recall letter was sent to customers in a e-mail beginning November 21, 2016 with a had copy sent via FedEx overnight with signature receipt beginning December 5, 2016. Devices will be picked up and replaced by Mallinckrodt. Questions can be directed to 1-877-KNOW-INO (1-877-566-9466).
Quantity
45
Distribution
US Distribution to the states of : CA, CO, DC, FL, GA, HI, IA, IL, IN, MI, MN, NC, NJ, NM, NY, OH, PA, and TX
Lot, serial or model codes
Serial Numbers: DS20070030, DS20070043, DS20070094, DS20070268, DS20090096, DS20090458, DS20090484, DS20090539, DS20090620, DS20090645, DS20100071, DS20100118, DS20100216, DS20100266, DS20100268, DS20100330, DS20100439, DS20100496, DS20100604, DS20100675, DS20100744, DS20100762, DS20100950, DS20101199, DS20101337, DS20101359, DS20101369, DS20101406, DS20101443, DS20101524, DS20110242, DS20110480, DS20110577, DS20110849, DS20110868, DS20110912, DS20110915, DS20110972, DS20110976, DS20110990, DS20120008, DS20120161, DS20120266, DS20120281, DS20120370.
510(k) numbers
["K131686"]
PMA numbers
not on file
Product code
MRN
Firm FEI
3004531588
Firm city
Madison
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register