Skip to the content

FDA Medical Device Recalls (openFDA)

MAST Biosurgery USA, Inc: The SurgiWrap Bioresorbable Adhesion Barrier Film. SurgiWrap 100x130x0.05mm (01) 18964050255027 (17) 170731(10)52804 LOT 52804 2017-08 REF25502-01 Product is wrapped in a foil pouch that is then sealed into an additional envelope.

Recall number
Z-1076-2015
Recalling firm
MAST Biosurgery USA, Inc
Product
The SurgiWrap Bioresorbable Adhesion Barrier Film. SurgiWrap 100x130x0.05mm (01) 18964050255027 (17) 170731(10)52804 LOT 52804 2017-08 REF25502-01 Product is wrapped in a foil pouch that is then sealed into an additional envelope.
Status
Terminated
Initiated
2014-08-26
Posted
2015-02-10
Terminated
2015-02-24
Reason
The expiration date of the bar code on the envelope label did not match the expiration date next to the hourglass. The printed expiration date was August of 2017, while the bar code date read as July 31, 2017. The correct expiration date was July 31, 2017.
Root cause
Labeling Change Control
Action
MAST Biosurgery labeled 49 new envelopes and sent them to the distributor, and subsequently had the distributor remove the contents of the outer envelope (instructions for use and the foil pouches containing the product) and place them into the newly supplied and properly labeled outer envelopes provided by them. The distributor requested to send back the original 49 enveloped. The second batch of new labels was inspected and attached to the new outer envelope to create a properly labeled outer envelope prior to shipping. The firm requested that once the distributor replaced the product and the IFU into the newly labeled envelopes that they provide documentation that the product had been repackaged into the new envelopes. The distributor was also requested to verify that they did not ship any product to customers with the incorrect label. The firm had the distributor return the 49 improperly with the incorrect labels and the first set of "corrected" labels to them for destruction. Further questions call (858) 550-8050.
Quantity
49 pieces
Distribution
Worldwide Distribution to the country of China.
Lot, serial or model codes
LOT 52804 2017-08
510(k) numbers
["K061473"]
PMA numbers
not on file
Product code
FTL
Firm FEI
3004661493
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register