FDA Medical Device Recalls (openFDA)

Iridex Corporation: Iridex TruFocus LIO Premiere" Laser Indirect Ophthalmoscope ( LIO ) The LIO is a headmounted indirect ophthalmoscope which is used with a handheld lens to view and perform laser treatments on a patient s retina. It allows binocular visualization of the peripheral retina and can be used to evaluate and treat patients.

Recall number
Z-1075-2018
Recalling firm
Iridex Corporation
Product
Iridex TruFocus LIO Premiere" Laser Indirect Ophthalmoscope ( LIO ) The LIO is a headmounted indirect ophthalmoscope which is used with a handheld lens to view and perform laser treatments on a patient s retina. It allows binocular visualization of the peripheral retina and can be used to evaluate and treat patients.
Status
Terminated
Initiated
2018-02-27
Posted
not on file
Terminated
2023-09-22
Reason
It was discovered that the laser energy was not going where intended to put it, resulting in focal cataracts and iris burns.
Root cause
Device Design
Action
Action Taken by IRIDEX: 1. A investigation was started immediately upon receipt of the returned devices from the Sloan Kettering complaint on November 3, 2017. CAPA 0125 was initiated to document the events, root cause, correction, corrective and preventive action and effectiveness of these actions. 2. A ship hold was initiated on November 15, 2017. 3.As described below, IRIDEX plans to notify customers and distributors that it is voluntarily recalling all models of the TruFocus LIO Premiere. 4. Public Warning A press release was prepared and sent out on February 23, 2018. For further questions please call (650) 962-8848 Ext. 3672
Quantity
US - 58 OUS - 46
Distribution
Worldwide Distribution - USA Distribution and to the countries of : China, Germany, Australia, Canada, Ireland, Lebanon, Japan, Israel and Qatar
Lot, serial or model codes
Device Identity : 87300, 87301, 87302, 87303, 87304
510(k) numbers
["K170718"]
PMA numbers
not on file
Product code
GEX
Firm FEI
2939653
Firm city
Mountain View
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register