Ortho-Clinical Diagnostics: VITROS 5,1 FS Chemistry System, Catalog Number 6801375, Unique Device Identifier Number 10758750001132; and VITROS 5,1 FS Chemistry System (Refurbished), Catalog Number 6801890, Unique Device Identifier Number 10758750001644; IVD. Product Usage: The VITROS 5,1 FS Chemistry System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.)

Recall number
Z-1075-2016
Recalling firm
Ortho-Clinical Diagnostics
Product
VITROS 5,1 FS Chemistry System, Catalog Number 6801375, Unique Device Identifier Number 10758750001132; and VITROS 5,1 FS Chemistry System (Refurbished), Catalog Number 6801890, Unique Device Identifier Number 10758750001644; IVD. Product Usage: The VITROS 5,1 FS Chemistry System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.)
Status
Terminated
Initiated
2015-12-21
Posted
2016-03-08
Terminated
2018-05-30
Reason
Increased U90-382 or 6LU condition codes generated by VITROS 250, 350, 5,1 FS, 4600 and 5600 Chemistry Systems when using Calibrator Kit 9, Lot 954. A trend of complaints regarding customers actions following U90-382 or 6LU condition codes was noted (See RES 72289 - VITROS Calibrator Kit 9, lot 954 recall).
Root cause
Error in labeling
Action
Ortho Clinical Diagnostics sent an URGENT PRODUCT CORRECTION NOTIFICATION Letter (Ref. CL2015-201, dated December 21, 2015) via FedEx overnight courier and/or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (only to those consignees who provided a PO Box as an address). The notification letter was sent to all customers who received the affected product. Foreign affiliates were informed by e-mail on 12/20/2015 of the issue and instructed to notify their consignees of the issue and required actions. The letter identified the affected product, problem and actions to be taken. For questions contact Customer Technical Services at 1-800-421-3311.
Quantity
US = 994; Foreign = 1405
Distribution
Worldwide Distribution - US Nationwide including Puerto Rico, and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela
Lot, serial or model codes
Software Version 2.8 and below; VITROS 5,1 FS Chemistry System Serial Numbers: J34000119 - J34002323; VITROS 5,1 FS Chemistry System (Refurbished) Serial Numbers: 34000133, 34000193, 34000197, 34000212,  34000285, 34000316, 34000318, 34000337, 34000360, 34000420, 34000451, 34000478, 34000495, 34000497, 34000529, 34000538, 34000552, 34000557,  34000590, 34000596, 34000601, 34000631, 34000642, 34000761, 34000799,  34000805, 34000833, 34000851, 34000893, 34000946, 34000947, 34001006,  34001172, 34001218, 34001240, 34001243, 34001248, 34001251, 34001274,  34001297, 34001317, 34001345, 34001378, 34001379, 34001389, 34001390, 34001396, 34001422, 34001472, 34001504, 34001569, 34001572, 34001612, 34001626, 34001633, 34001668, 34001755, 34001809, 34001840, 34001850,  34001875, 34001897, 34000123, 34000143, 34000191, 34000215, 34000219,  34000235, 34000251, 34000256, 34000260, 34000261, 34000319, 34000331, 34000380, 34000382, 34000406, 34000408, 34000412, 34000426, 34000431,  34000454, 34000464, 34000473, 34000501, 34000523, 34000530, 34000541,  34000542, 34000546, 34000548, 34000553, 34000558, 34000576, 34000577, 34000584, 34000585, 34000609, 34000612, 34000615, 34000616, 34000617,  34000618, 34000620, 34000625, 34000637, 34000644, 34000672, 34000744,  34000745, 34000755, 34000759, 34000767, 34000776, 34000793, 34000800, 34000814, 34000831, 34000844, 34000846, 34000852, 34000856, 34000870,  34000876, 34000887, 34000892, 34000902, 34000912, 34000920, 34000938,  34000956, 34000969, 34000976, 34001007, 34001009, 34001044, 34001045,  34001052, 34001071, 34001075, 34001098, 34001099, 34001111, 34001129,  34001154, 34001161, 34001171, 34001182, 34001192, 34001195, 34001201,  34001204, 34001208, 34001212, 34001234, 34001247, 34001254, 34001271,  34001278, 34001284, 34001288, 34001289, 34001293, 34001298, 34001301,  34001302, 34001316, 34001325, 34001337, 34001340, 34001348, 34001367,  34001369, 34001399, 34001400, 34001407, 34001410, 34001416, 34001423,  34001427, 34001429, 34001457, 34001490, 34001518, 34001526, 34001551, 34001573, 34001580, 34001581, 34001582, 34001587, 34001589, 34001594,  34001610, 34001618, 34001623, 34001627, 34001673, 34001688, 34001693,  34001725, 34001738, 34001745, 34001747, 34001790, 34001796, 34001798,  34001810, 34001822, 34001831, 34001833, 34001849, 34001855, 34001859,  34001860, 34001867, 34001871, 34001872, 34001881, 34001905, 34001916,  34001923, 34001995
510(k) numbers
["K031924","K063144"]
PMA numbers
not on file
Product code
JJE
Firm FEI
1000136573
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register