FDA Medical Device Recalls (openFDA)

Radiometer America Inc: pCO2 D788 Membranes. Distributed by Radiometer America, Inc., Westlake, OH. An automatic or semi-automatic instrument used to identify and quantify two or more gases, and sometimes electrolytes, in whole blood using multiple special electrodes.

Recall number
Z-1075-2014
Recalling firm
Radiometer America Inc
Product
pCO2 D788 Membranes. Distributed by Radiometer America, Inc., Westlake, OH. An automatic or semi-automatic instrument used to identify and quantify two or more gases, and sometimes electrolytes, in whole blood using multiple special electrodes.
Status
Terminated
Initiated
2013-07-05
Posted
2014-02-24
Terminated
2015-09-19
Reason
RADIOMETER has become aware that some D788 pCO2 membranes can cause biased measurement results on patient results as well as QC results.
Root cause
Under Investigation by firm
Action
On 7/18/2013, the firm sent Product Notification letters to their customers.
Quantity
38,890 units (7,860 units imported into the US)
Distribution
Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV & WY, and the countries of Bermuda and Canada.
Lot, serial or model codes
Model #: 942-063 Lot #'s: R338 to R493
510(k) numbers
["K041874","K980130"]
PMA numbers
not on file
Product code
CHL
Firm FEI
1523456
Firm city
Westlake
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register