FDA Medical Device Recalls (openFDA)
Cardinal Health: AirLife nCPAP System Driver; a nasal continuous positive airway pressure device, which is part of a larger system that includes an nCPAP generator, fixation device and patient interface (prongs and.or mask); Cardinal Health, McGaw Park, IL 60085; cat. 006900
- Recall number
- Z-1075-2007
- Recalling firm
- Cardinal Health
- Product
- AirLife nCPAP System Driver; a nasal continuous positive airway pressure device, which is part of a larger system that includes an nCPAP generator, fixation device and patient interface (prongs and.or mask); Cardinal Health, McGaw Park, IL 60085; cat. 006900
- Status
- Terminated
- Initiated
- 2007-06-01
- Posted
- 2007-07-12
- Terminated
- 2007-12-12
- Reason
- Under isolated conditions, the device may experience issues with O2 Calibration, which may result in either a nuisance alarm or a delivered FiO2 of up to 5% below the specified tolerance.
- Root cause
- Other
- Action
- Cardinal Health sent the accounts recall letters dated 6/1/07 via UPS next day delivery, which listed the catalog number and affected serial numbers of the CPAP Driver, and informed them that under certain isolated conditions, the CPAP Driver may experience issues with O2 Calibration, which may result in either a nuisance alarm or a delivered FiO2 of up to 5% below the specified tolerance. Customers have experienced this under the following conditions: 1) air inlet pressure is greater than O2 pressure by 5 to 10 psi and 2) FiO2 settings less than 45%. Separate recall letters were sent to those accounts who had filed complaints of O2 Calibration Errors, stating that their device will be replaced by the end of June, 2007; those accounts who purchased the devices, but had not reported O2 Calibration Errors, stating that a Cardinal Health representative will upgrade the firmware on the devices at their facility by the end of August, 2007; and those accounts who had had demonstration units in their possession, advising them to notify their sales representative if any of the conditions mentioned in the letter occurred. The accounts were requested to complete and return via fax at 847-689-9101, the enclosed acknowledgment form, indicating the number of affected drivers in their possession, and were advised that they would be contacted by a Cardinal Health representative to arrange for the replacement/service of their drivers. Any questions were directed to Quality Systems-Professional Services at 1-800-292-9332 or 1-800-327-7503.
- Quantity
- 48 units
- Distribution
- Nationwide.
- Lot, serial or model codes
- Catalog Number 006900, Firmware Version 2.02 with the following Serial Numbers: 0107DL002, 0107DL003, 0107DL004, 0107DL011, 0107DL008, 0606DL001, 0606DL002, 0606DL009, 0606DL012, 0606DL013, 0606DL014, 0606DL015, 0606DL017, 0606DL018, 0606DL019, 0606DL021, 0606DL022, 0606DL023, 0606DL024, 0606DL025, 0606DL026, 0606DL027, 0606DL033, 0606DL040, 1106DL001, 1106DL002, 1106DL003, 1106DL004, 1106DL005, 1106DL006, 1106DL007, 1106DL008, 1106DL009, 1106DL010, 1106DL011, 1106DL012, 1106DL013, 1106DL014, 1106DL015, 1106DL016, 1106DL017, 1106DL018, 1106DL019, 1106DL020, 1206DL001, 1206DL002, 1206DL003, 1206DL004, 1206DL005, 1206DL006, 1206DL007, 1206DL008, 1206DL009, 1206DL010, 1206DL011, 1206DL012, 1206DL013, 1206DL014, 1206DL015, 1206DL017, 1206DL018, 1206DL019, 1206DL020, 1206DL022, 1206DL023, 1206DL025, 1206DL016
- 510(k) numbers
- ["K051226"]
- PMA numbers
- not on file
- Product code
- BZD
- Firm FEI
- 3001236905
- Firm city
- Park
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28