Skip to the content

FDA Medical Device Recalls (openFDA)

Getinge Usa Sales Inc: Vaporizer Sevoflurane, Quik-Fil. Model Number: 6682285. Device is a component of Flow-i C20, Flow-i C30, Flow-i C40, Flow-c, Flow-e. For use during anesthesia.

Recall number
Z-1074-2025
Recalling firm
Getinge Usa Sales Inc
Product
Vaporizer Sevoflurane, Quik-Fil. Model Number: 6682285. Device is a component of Flow-i C20, Flow-i C30, Flow-i C40, Flow-c, Flow-e. For use during anesthesia.
Status
Open, Classified
Initiated
2025-01-15
Posted
2025-02-07
Terminated
not on file
Reason
Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which in worst case may lead to lung edema and/or severe hypocalcemia, blistering, superficial ulceration, and/or hypomagnesemia.
Root cause
Device Design
Action
Getinge notified consignees on 01/15/2024 via URGENT MEDICAL DEVICE RECALL letter. Consignees were provided additional information which is complementary to the existing user manual and were instructed to ensure all affected product inventory has been quarantined, consignees are entitled to credit/replacement, the recall information is forwarded to affected personnel and any customers if it was further distributed and complete and return the provided URGENT MEDICAL DEVICE RECALL RESPONSE FORM.
Quantity
6 units
Distribution
Domestic distribution to Kentucky and Connecticut.
Lot, serial or model codes
Model Number: 6682285. UDI-DI: 07325710001141. Serial Numbers: 3761, 7635, 7633, 7630, 7632, 7631.
510(k) numbers
["K191027"]
PMA numbers
not on file
Product code
BSZ
Firm FEI
3013876692
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28