FDA Medical Device Recalls (openFDA)
Getinge Usa Sales Inc: Vaporizer Sevoflurane, Quik-Fil. Model Number: 6682285. Device is a component of Flow-i C20, Flow-i C30, Flow-i C40, Flow-c, Flow-e. For use during anesthesia.
- Recall number
- Z-1074-2025
- Recalling firm
- Getinge Usa Sales Inc
- Product
- Vaporizer Sevoflurane, Quik-Fil. Model Number: 6682285. Device is a component of Flow-i C20, Flow-i C30, Flow-i C40, Flow-c, Flow-e. For use during anesthesia.
- Status
- Open, Classified
- Initiated
- 2025-01-15
- Posted
- 2025-02-07
- Terminated
- not on file
- Reason
- Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which in worst case may lead to lung edema and/or severe hypocalcemia, blistering, superficial ulceration, and/or hypomagnesemia.
- Root cause
- Device Design
- Action
- Getinge notified consignees on 01/15/2024 via URGENT MEDICAL DEVICE RECALL letter. Consignees were provided additional information which is complementary to the existing user manual and were instructed to ensure all affected product inventory has been quarantined, consignees are entitled to credit/replacement, the recall information is forwarded to affected personnel and any customers if it was further distributed and complete and return the provided URGENT MEDICAL DEVICE RECALL RESPONSE FORM.
- Quantity
- 6 units
- Distribution
- Domestic distribution to Kentucky and Connecticut.
- Lot, serial or model codes
- Model Number: 6682285. UDI-DI: 07325710001141. Serial Numbers: 3761, 7635, 7633, 7630, 7632, 7631.
- 510(k) numbers
- ["K191027"]
- PMA numbers
- not on file
- Product code
- BSZ
- Firm FEI
- 3013876692
- Firm city
- Wayne
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28