FDA Medical Device Recalls (openFDA)
Smiths Medical ASD Inc.: Power PORT-A-CATH II, TRAY POWER PAC II. Labeled as: a. VEN, PU, 2.6MM, UL 8.5 FR/INTRO, VALVED (1/EA); b. VEN LP, PU, 2.6MM, UL 8.5 FR/INTRO, VALVED (1/EA); c. VEN LP, PU, 1.9MM, UL, 6 FR/INTRO, VALVED (1/EA); d. DL, LP, PU, 2.2MM, WING-LOCK, 7FR, INTRO (1/EA); e. DL, LP, PU, 3.2MM, WING-LOCK, 10 FR/INTRO (1/EA); f. DL, LP, PU, 2.2MM, WING-LOCK, 7FR/INTRO, VALVED (1/EA); g. DL, LP, PU, 3.2MM, WING-LOCK, 10 FR/INTRO, VALVED (1/EA)
- Recall number
- Z-1074-2018
- Recalling firm
- Smiths Medical ASD Inc.
- Product
- Power PORT-A-CATH II, TRAY POWER PAC II. Labeled as: a. VEN, PU, 2.6MM, UL 8.5 FR/INTRO, VALVED (1/EA); b. VEN LP, PU, 2.6MM, UL 8.5 FR/INTRO, VALVED (1/EA); c. VEN LP, PU, 1.9MM, UL, 6 FR/INTRO, VALVED (1/EA); d. DL, LP, PU, 2.2MM, WING-LOCK, 7FR, INTRO (1/EA); e. DL, LP, PU, 3.2MM, WING-LOCK, 10 FR/INTRO (1/EA); f. DL, LP, PU, 2.2MM, WING-LOCK, 7FR/INTRO, VALVED (1/EA); g. DL, LP, PU, 3.2MM, WING-LOCK, 10 FR/INTRO, VALVED (1/EA)
- Status
- Terminated
- Initiated
- 2017-12-04
- Posted
- 2018-03-15
- Terminated
- 2020-06-12
- Reason
- Certain models and lots of PORT A CATH implantable port kits may have been misassembled. The kits may incorrectly contain a 6Fr or 6.5Fr size introducer instead of the correct 8.5Fr size introducer. The Power PORT A CATH II kit may also contain the incorrect locking collars in addition to the incorrect sized introducer.
- Root cause
- Process design
- Action
- Customers were notified via letter on about 12/04/2017. Instructions include to examine current inventory for affected product, collect and isolate any potentially affected product, determine the total number of potentially affected product in possession and complete and return the Urgent Medical Device Recall Response Form, return any affected product on hand, and notify customers if the affected products were further distributed.
- Quantity
- 1,077 units total
- Distribution
- Worldwide Distribution. US nationwide including Puerto Rico, Austria, Bahamas, Belgium, Dominican Republic, Netherlands, Spain, and Turks and Caicos.
- Lot, serial or model codes
- a. Item No. 21-4474-24 (Lot No. 87X040, 87X117); b. Item No. 21-4475-24 (Lot No. 87X044); c. Item No. 21-4477-24 (Lot No. 87X097); d. Item No. 21-8466-24 (Lot No. 87X171); e. Item No. 21-8468-24 (Lot No. 87X172, 87X244, 87X180); f. Item No. 21-8469-24 (Lot No. 87X151); g. Item No. 21-8470-24 (Lot No. 87X173, 87X254, 87X194, 87X198, 87X237)
- 510(k) numbers
- ["K093414"]
- PMA numbers
- not on file
- Product code
- LJT
- Firm FEI
- 3012307300
- Firm city
- Minneapolis
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28