Skip to the content

FDA Medical Device Recalls (openFDA)

Smiths Medical ASD Inc.: Power PORT-A-CATH II, TRAY POWER PAC II. Labeled as: a. VEN, PU, 2.6MM, UL 8.5 FR/INTRO, VALVED (1/EA); b. VEN LP, PU, 2.6MM, UL 8.5 FR/INTRO, VALVED (1/EA); c. VEN LP, PU, 1.9MM, UL, 6 FR/INTRO, VALVED (1/EA); d. DL, LP, PU, 2.2MM, WING-LOCK, 7FR, INTRO (1/EA); e. DL, LP, PU, 3.2MM, WING-LOCK, 10 FR/INTRO (1/EA); f. DL, LP, PU, 2.2MM, WING-LOCK, 7FR/INTRO, VALVED (1/EA); g. DL, LP, PU, 3.2MM, WING-LOCK, 10 FR/INTRO, VALVED (1/EA)

Recall number
Z-1074-2018
Recalling firm
Smiths Medical ASD Inc.
Product
Power PORT-A-CATH II, TRAY POWER PAC II. Labeled as: a. VEN, PU, 2.6MM, UL 8.5 FR/INTRO, VALVED (1/EA); b. VEN LP, PU, 2.6MM, UL 8.5 FR/INTRO, VALVED (1/EA); c. VEN LP, PU, 1.9MM, UL, 6 FR/INTRO, VALVED (1/EA); d. DL, LP, PU, 2.2MM, WING-LOCK, 7FR, INTRO (1/EA); e. DL, LP, PU, 3.2MM, WING-LOCK, 10 FR/INTRO (1/EA); f. DL, LP, PU, 2.2MM, WING-LOCK, 7FR/INTRO, VALVED (1/EA); g. DL, LP, PU, 3.2MM, WING-LOCK, 10 FR/INTRO, VALVED (1/EA)
Status
Terminated
Initiated
2017-12-04
Posted
2018-03-15
Terminated
2020-06-12
Reason
Certain models and lots of PORT A CATH implantable port kits may have been misassembled. The kits may incorrectly contain a 6Fr or 6.5Fr size introducer instead of the correct 8.5Fr size introducer. The Power PORT A CATH II kit may also contain the incorrect locking collars in addition to the incorrect sized introducer.
Root cause
Process design
Action
Customers were notified via letter on about 12/04/2017. Instructions include to examine current inventory for affected product, collect and isolate any potentially affected product, determine the total number of potentially affected product in possession and complete and return the Urgent Medical Device Recall Response Form, return any affected product on hand, and notify customers if the affected products were further distributed.
Quantity
1,077 units total
Distribution
Worldwide Distribution. US nationwide including Puerto Rico, Austria, Bahamas, Belgium, Dominican Republic, Netherlands, Spain, and Turks and Caicos.
Lot, serial or model codes
a. Item No. 21-4474-24 (Lot No. 87X040, 87X117);  b. Item No. 21-4475-24 (Lot No. 87X044);  c. Item No. 21-4477-24 (Lot No. 87X097);  d. Item No. 21-8466-24 (Lot No. 87X171);  e. Item No. 21-8468-24 (Lot No. 87X172, 87X244, 87X180);  f. Item No. 21-8469-24 (Lot No. 87X151);  g. Item No. 21-8470-24 (Lot No. 87X173, 87X254, 87X194, 87X198, 87X237)
510(k) numbers
["K093414"]
PMA numbers
not on file
Product code
LJT
Firm FEI
3012307300
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register