FDA Medical Device Recalls (openFDA)
Lifescan Inc: All OneTouch Verio IQ Blood Glucose Meters sold as: Verio IQ System Kits - Meter; Verio IQ Starter kits - Meter; Verio IQ Warranty Meter. Product Usage: The OneTouch VerioIQ Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The system is intended to be used by a single patient and should not be used for testing multiple patients. The OneTouch VerioIQ Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control.
- Recall number
- Z-1074-2013
- Recalling firm
- Lifescan Inc
- Product
- All OneTouch Verio IQ Blood Glucose Meters sold as: Verio IQ System Kits - Meter; Verio IQ Starter kits - Meter; Verio IQ Warranty Meter. Product Usage: The OneTouch VerioIQ Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The system is intended to be used by a single patient and should not be used for testing multiple patients. The OneTouch VerioIQ Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control.
- Status
- Terminated
- Initiated
- 2013-03-11
- Posted
- 2013-04-12
- Terminated
- 2015-08-03
- Reason
- The Verio IQ meter will shut off and revert to set up mode at glucose values above 1023 instead of displaying EXTREME HIGH GLUCOSE.
- Root cause
- Software design
- Action
- Lifescan sent an Urgent Medical Device Voluntary Recall letter dated March 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to identify and hold all affected product in inventory and discontinue distributing. The letter states that a Lifescan sales representative will contact customers to arrange for pick up and replacement.
- Quantity
- US 710,413
- Distribution
- USA Nationwide Distribution
- Lot, serial or model codes
- Part numbers: Verio IQ Systems Kits: 022 267-01, 022 267-01, 022 267-03; Verio IQ Starter Kits: 022 268 01, 022 268-02; Verio IQ Warranty Kits: 022 275 01, 022 275 02. All lots and serial numbers
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- NBW
- Firm FEI
- 2939301
- Firm city
- Milpitas
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28