Skip to the content

FDA Medical Device Recalls (openFDA)

Argon Medical Devices, Inc: Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 19, 1/19 GA Flexcore Biopsy Needle, 7F Introducer Sheath W/Curved Metal Stiffener, Tissue Removal Swabs, 5F Straight Catheter, 5F Curved Catheter, Sterile EO

Recall number
Z-1073-2021
Recalling firm
Argon Medical Devices, Inc
Product
Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 19, 1/19 GA Flexcore Biopsy Needle, 7F Introducer Sheath W/Curved Metal Stiffener, Tissue Removal Swabs, 5F Straight Catheter, 5F Curved Catheter, Sterile EO,
Status
Terminated
Initiated
2020-12-10
Posted
not on file
Terminated
2023-01-31
Reason
Due to a manufacturing error their is a potential that the sheath within the biopsy kit can separate during use if the tip is flexed or experiences bending stresses. This could result in need for surgical intervention to preclude or reverse permanent damage to a body structure or permanent impairment of a body function.
Root cause
Process change control
Action
On 12/11/2020, Argon Medical Devices, Inc. distributed an "Urgent-Product Recall Notice" to all affected consignees via US mail. In addition to informing consignees about the recall device, the firm ask consignees to take the following actions: 1. Complete the response form and return it to us as quickly as possible. This will allow us to begin staging replacement product for you and minimize interruption to service. 2. All affected product should be returned to our Argon Athens facility using RGA#______, attention Arbee Cummings. The mailing address is as follows: RGA# ____, Argon Medical Devices, Inc.1445 Flat Creek Road, Athens, TX 75751 USA 3. Argon Medical will ship your replacement devices as soon as we receive your returned product. Your assistance in accounting for the affected devices in your possession is greatly appreciated. If you have any questions about this letter or about the recall action it describes, please contact Beckie.Ellis@argonmedical.com. You may also contact Ms. Arbee Cummings at Arbee.Cummings@argonmedical.com or Ms. Andrea Wieczor at Andrea.Wieczor@argonmedical.com.
Quantity
863 unit
Distribution
US: AL AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA,, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX,, UT, VA, WA, WI, WV, WY OUS: Switzerland, Saudi Arabia, Hong Kong, Thailand, Portugal, Taiwan, Austria, Brazil, Netherlands, Italy, Germany, Colombia, Singapore, Canada, View Nam, Slovakia, Spain
Lot, serial or model codes
Lot Number: 1428364, 1451428, 1486448
510(k) numbers
["K022634"]
PMA numbers
not on file
Product code
DYB
Firm FEI
1625425
Firm city
Athens
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register