Skip to the content

FDA Medical Device Recalls (openFDA)

NxStage Medical, Inc.: NxStage PureFlow-B Solution; Model/Catalog Number: RFP-401. Product is packaged in a case containing two individual 5L bags. The NxStage PureFlow-B Solution is indicated for use with renal replacement therapy systems that utilize sterile premixed dialysate during hemodialysis.

Recall number
Z-1073-2014
Recalling firm
NxStage Medical, Inc.
Product
NxStage PureFlow-B Solution; Model/Catalog Number: RFP-401. Product is packaged in a case containing two individual 5L bags. The NxStage PureFlow-B Solution is indicated for use with renal replacement therapy systems that utilize sterile premixed dialysate during hemodialysis.
Status
Terminated
Initiated
2014-01-24
Posted
2014-02-24
Terminated
2014-08-29
Reason
Mislabeled product.
Root cause
Packaging process control
Action
NxStage sent a Voluntary Medical Device Recall letter dated January 24, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. They were asked to check all boxes in inventory for the recalled lot and segregate and quarantine them. In addition, they were asked to complete the recall reply form and return it by fax to NxStage Medical as well as to contact NxStage Customer service to arrange for return of all affected product and for replacement to be sent. Distributor was asked to notify all downstream customers and to provide NxStage Medical with a list of customers. Please complete and fax this form to 978-687-4810 even if you do not have any of the affected product on hand. If you have any questions or comments, please feel free to contact NxStage Customer Service at 1-866- NXSTAGE (1-866-697-8243).
Quantity
969 cases (2 bags per case)
Distribution
US Distribution including the states of NY, TX, ID, AR, CA and AL
Lot, serial or model codes
Lot Number: F113662
510(k) numbers
["K053286"]
PMA numbers
not on file
Product code
KPO
Firm FEI
3003464075
Firm city
Lawrence
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register