FDA Medical Device Recalls (openFDA)
Gyrus ACMI Corporation: Gyrus CyberWand Ultrasonic Lithotripter System includes: CW-USLS CyberWand USL System CW-USLT CyberWand USL Transducer CW-USLRBP CyberWand USL Renal/Bladder Probe Set CW-USLTT CyberWand USL Transducer Sterilization Tray
- Recall number
- Z-1073-2007
- Recalling firm
- Gyrus ACMI Corporation
- Product
- Gyrus CyberWand Ultrasonic Lithotripter System includes: CW-USLS CyberWand USL System CW-USLT CyberWand USL Transducer CW-USLRBP CyberWand USL Renal/Bladder Probe Set CW-USLTT CyberWand USL Transducer Sterilization Tray
- Status
- Terminated
- Initiated
- 2007-05-23
- Posted
- 2007-08-09
- Terminated
- 2010-05-25
- Reason
- Sterilization Instructions not adequate for sterilizing.
- Root cause
- Other
- Action
- Gyrus has informed their sales force to visit each customer that has a CyberWand system(s) and hand deliver a letter, dated May 23, 2007, that explains the situation and requests that the customer cease using the unit and segregate it to prevent accidental use. The letter also explains that the firm is collecting any sterilization trays with printed sterilization instructions and replacing them. The rep will assist the customer in complying. The customer will also be required to complete a reply form to track compliance.
- Quantity
- 53 units
- Distribution
- Nationwide and Canada.
- Lot, serial or model codes
- Product distributed prior to May 7, 2007.
- 510(k) numbers
- ["K052135"]
- PMA numbers
- not on file
- Product code
- FFK
- Firm FEI
- 3003790304
- Firm city
- Southborough
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28