LivaNova USA, Inc.: SenTiva, REF: 1000; SenTiva Duo, REF: 1000-D; Generators Part of the VNS Therapy System. Used for Vagus Nerve Stimulation (VNS).

Recall number
Z-1072-2025
Recalling firm
LivaNova USA, Inc.
Product
SenTiva, REF: 1000; SenTiva Duo, REF: 1000-D; Generators Part of the VNS Therapy System. Used for Vagus Nerve Stimulation (VNS).
Status
Open, Classified
Initiated
2024-12-12
Posted
2025-01-31
Terminated
not on file
Reason
Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.
Root cause
Device Design
Action
On 12/12/2024, correction notices were mailed to customers asking them to do the following: 1. Monitor patient frequently and perform an on-demand System Diagnostic test at each visit per labeling (typical patient visit cadence is every 3-12 months). See VNS Therapy Physician's Manual, found in the Manuals Section https://www.livanova.com/epilepsy-vnstherapy/en-us/hcp/product-manuals or https://www.livanova.com/depression/en-us/healthcare-provider-resources. 2. Counsel prospective and current patients to ensure that they are aware of and understand the risks. You may elect to provide the patient letter. 3. Contact Customer Quality at 1-866-882-8804 or e-mail at cservices@livanova.com if communication with the generator using the Programming System indicates that: i. Diagnostic tests cannot be performed (Error Code 254 - Test Interrupted error message) ii. System Diagnostic test displays a LOW lead impedance result. 4. Encourage patients to: i. Notify you if there is a change in perceived clinical symptoms (e.g. increase in seizures or depressive symptoms, etc.). ii. Notify you if they no longer perceive any form of stimulation. 5. Ensure that this notice is communicated to all personnel within your organization who need to be aware. 6. Complete and return the response form via email to M1000FieldAction@livanova.com Firm is updating the VNS Therapy Instructions for Use (IFU) to reflect the patient management recommendations relevant to this issue. For questions contact the firm's Customer Quality at 1-866-882-8804 or e-mail cservices@livanova.com On 6/16/2025, updated recall notices were delivered in person notifying customers that devices with a design change were available so affected generators remaining in inventory should be replaced. New devices with the design change have serial numbers greater than 500,000. There are no new actions for devices that have already been implanted.
Quantity
81,606
Distribution
US: AK, AZ, CA, CO, CT, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, PN, SC, SD, TN, TX, VA, VT, WA, WI, WV, UT. OUS: AE, AR, AT, BE, BG, BR, CA, CL, CO, CR, CZ, DE, DK, GR, HR, IL, IN, IT, JO, KW, LB, MK, NL, OM, PA, PE, PL, QA, RO, RS, SA, SE, SG, SI, TR, TT, UK, ZA.
Lot, serial or model codes
REF/UDI-DI: 1000/05425025750405, 1000-D/05425025750528 All Serial Numbers less than 500,000
510(k) numbers
not on file
PMA numbers
["P970003"]
Product code
LYJ
Firm FEI
1000117004
Firm city
Houston
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register