FDA Medical Device Recalls (openFDA)
Ortho-Clinical Diagnostics: VITROS 3600 Immunodiagnostics Systems, Catalog # 6802783; COMMON/USUAL NAME: VITROS 3600 --- NOTE: VITROS 3600 that are e-Connected are not affected by this issue. Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents.
- Recall number
- Z-1072-2012
- Recalling firm
- Ortho-Clinical Diagnostics
- Product
- VITROS 3600 Immunodiagnostics Systems, Catalog # 6802783; COMMON/USUAL NAME: VITROS 3600 --- NOTE: VITROS 3600 that are e-Connected are not affected by this issue. Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents.
- Status
- Terminated
- Initiated
- 2011-04-13
- Posted
- 2012-02-21
- Terminated
- 2016-05-09
- Reason
- Ortho Clinical Diagnostics (OCD) received customer complaints regarding the loss of calibration data following the installation of a development Assay Data Disk (ADD) that was available in the on-screen mailbox on the Main Menu System Status screen of their VITROS 5600 Integrated System. OCD's investigation determined that ADD, Data Release Version (DRV) 9241 or 9251 was electronically transferr
- Root cause
- Software design
- Action
- Ortho-Clinical Diagnostics (OCD) sent an Important Product Correction Notification letters dated April 13, 2011 via FedEx overnight to customers with an affected VITROS 3600 or 5600 System (those not e-Connected). Foreign affiliate consignees were informed of this action via e-mail on April 13, 2011. The letters described the affected products, problem and actions to be followed. The letters states that an OCD Field Engineer will contact customers who has the affected products and perform a Modification (MOD) procedure that will remove ADD DRV 9241 or 9251 from the on-screen mailbox of their system(s). Customers were instructed to complete and return the accompanying Confirmation of Receipt Form and fax to1-888-557-3759 or 1-585-453-4110. The letter asks that distributors forward a copy of the enclosed customer communication to all customers who were shipped an affected VITROS 3600 or 5600 System. For questions, please contact OCD's Customer Technical Services at 1-800-421-3311.
- Quantity
- Domestic - 15 units, foreign - 81 units
- Distribution
- Worldwide Distribution - USA (nationwide) including the countries of: Brazil, Canada, Chile, China, England, France, Germany, India, Italy, Japan, Mexico, Singapore, Spain and Venezuela.
- Lot, serial or model codes
- All VITROS¿ 3600 Systems Catalog # 6802783 that have not been eConnected which may have Assay Data Disk (ADD) Data Release Version (DRV) 9241 or 9251. J/Serial Numbers: 36000206, 36000268, 36000269, 36000290, 36000319, 36000404, 36000414, 36000416, 36000426, 36000430, 36000453, 36000464, 36000466, 36000467, 36000475, 36000143, 36000151, 36000152, 36000159, 36000165, 36000187, 36000194, 36000198, 36000204, 36000222, 36000228, 36000257, 36000258, 36000271, 36000275, 36000279, 36000301, 36000306, 36000307, 36000313, 36000315, 36000322, 36000324, 36000325, 36000326, 36000327, 36000330, 36000332, 36000333, 36000335, 36000336, 36000338, 36000342, 36000347, 36000348, 36000352, 36000353, 36000355, 36000360, 36000364, 36000365, 36000372, 36000374, 36000382, 36000384, 36000389, 36000390, 36000391, 36000392, 36000394, 36000395, 36000400, 36000401, 36000403, 36000405, 36000408, 36000421, 36000422, 36000425, 36000429, 36000431, 36000434, 36000438, 36000441, 36000442, 36000443, 36000448, 36000461, 36000462, 36000468, 36000469, 36000470, 36000471, 36000472, 36000473, 36000476, 36000477, 36000478, 36000479, 36000481, 36000483. -- NOTE: VITROS¿ 3600 that are e-Connected are not affected by this issue.
- 510(k) numbers
- ["K083173"]
- PMA numbers
- not on file
- Product code
- JJE
- Firm FEI
- 1000136573
- Firm city
- Rochester
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28