J T Posey Company: Posey Biothane Restraint Cuffs, Catalog Number 2900. The device is intended to limit a patient's movement to the extent necessary for treatment or protection of the patient or others.

Recall number
Z-1071-2013
Recalling firm
J T Posey Company
Product
Posey Biothane Restraint Cuffs, Catalog Number 2900. The device is intended to limit a patient's movement to the extent necessary for treatment or protection of the patient or others.
Status
Terminated
Initiated
2012-10-31
Posted
2013-04-09
Terminated
2014-07-17
Reason
The firm voluntarily recalled Biothane Restraint Cuffs, Catalog No. 2900, that did not meet the required specifications for tensile strength. Failure of these rivets could allow the cuff to become disconnected from the connecting strap and allow patients to regain free use of the limb.
Root cause
Process design
Action
J. T. Posey sent an Medical Device Recall (Removal) letter to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter instructed the customer that Posey Company is voluntarily recalling all units of Posey BiothaneQD Cuffs, Catalog Number 2900. All lots of this product are being recalled. Reason for recall was listed:"Posey has identified rivets on the Biothane~ Restraint Cuffs, Catalog No. 2900, that did not meet the required specifications for tensile strength. Failure of these rivets could allow the cuff to become disconnected from the connecting strap and allow patients to regain free use of the limb. No patient injuries or deaths have reported." Please immediately discontinue further distribution or use of Posey Biothane~ Cuffs, Catalog Number 2900. Please Complete and return the Acknowledgement and Receipt Form to Posey within 7 days using the provided self-addressed, stamped envelope, by fax to 626-443-5418 or via e-mail to RegulatoryAffairs@Posey.com. Return this form even if you do not have any devices in your facility. Please return directly to Posey ill! units you have of Biothane~ Cuffs, Catalog Number 2900, using the enclosed pre-paid shipping label containing an RMA number. Posey will provide replacement product or credit upon receipt of the product. If you have any further questions please call (626) 443-3143.
Quantity
1673 units
Distribution
Nationwide Distribution
Lot, serial or model codes
The model/catalog number for the device is 2900. All lots are being recalled.
510(k) numbers
not on file
PMA numbers
not on file
Product code
FMQ
Firm FEI
3016004577
Firm city
Arcadia
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register