FDA Medical Device Recalls (openFDA)
Boston Scientific Corp: Convoy Advanced Delivery Sheath Kit, Model Numbers: M0045660ST0, M0045661ST0, M0045662ST0, M0045663ST0, M0045665ST0, M0045667ST0, M0045670ST0, M0045672ST0, M0045675ST0, M0045676ST0, M0045690ST0, M0045890ST0, M0045748ST0 and M0045749ST0. Manufactured by Thomas Medical Products, Malvern, PA. Product is intended for use to facilitate the intracardiac placement of interventional devices.
- Recall number
- Z-1071-2009
- Recalling firm
- Boston Scientific Corp
- Product
- Convoy Advanced Delivery Sheath Kit, Model Numbers: M0045660ST0, M0045661ST0, M0045662ST0, M0045663ST0, M0045665ST0, M0045667ST0, M0045670ST0, M0045672ST0, M0045675ST0, M0045676ST0, M0045690ST0, M0045890ST0, M0045748ST0 and M0045749ST0. Manufactured by Thomas Medical Products, Malvern, PA. Product is intended for use to facilitate the intracardiac placement of interventional devices.
- Status
- Terminated
- Initiated
- 2008-12-24
- Posted
- 2009-04-09
- Terminated
- 2010-04-06
- Reason
- The marker band may detach from the sheath and a clinical failure could result in an embolism.
- Root cause
- Process design
- Action
- The firm initiated the recall on December 24, 2009 and notified consignees via "Urgent Company Initiated Medical Device Recall - Immediate Action Required" letters which explain the reason for the recall and request the affected product be returned to the Boston Scientific Corporation Distribution Center. An amendment letter with two additional Convoy Advanced Delivery Sheath Kit models (M0045748ST0 and M0045749ST0) was sent to affected customers on January 27, 2009. Direct questions about this recall to your local Sales Representative.
- Quantity
- 31,340 of all units
- Distribution
- Worldwide Distribution.
- Lot, serial or model codes
- Batch Numbers: S1721 4 S1899 6 S2103 9 S2287 6 S2497 2 S1724 9 S1903 2 S2105 6 S2287 7 S2513 9 S1732 2 S1913 1 S2109 5 S2295 6 S2522 7 S1733 8 S1913 2 S2115 2 S2295 9 S2524 0 S1735 6 S1913 3 S2118 9 S2296 4 S2536 0 S1737 1 S1926 5 S2123 4 S2297 1 S2553 7 S1737 2 S1930 1 S2135 6 S2299 4 S2553 8 S1737 4 S1956 3 S2151 1 S2299 5 S2555 1 S1757 4 S1962 0 S2156 4 S2301 4 S2557 8 S1766 6 S1962 1 S2161 5 S2305 1 S2562 9 S1766 7 S1964 4 S2166 4 S2306 9 S2566 1 S1766 8 S1970 4 S2166 5 S2307 0 S2566 4 S1792 6 S2015 1 S2166 6 S2308 9 S2567 6 S1792 S2015 S2180 S2309 and S2579.
- 510(k) numbers
- ["K013866"]
- PMA numbers
- not on file
- Product code
- DYB
- Firm FEI
- 3001236349
- Firm city
- San Jose
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28