Skip to the content

FDA Medical Device Recalls (openFDA)

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Allura Xper FD20; Model Number: 722012;

Recall number
Z-1070-2026
Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product
Allura Xper FD20; Model Number: 722012;
Status
Open, Classified
Initiated
2025-12-15
Posted
2026-01-13
Terminated
not on file
Reason
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Root cause
Other
Action
On December 15, 2025, URGENT Medical Device Correction letters were sent to customers. Actions to be taken: a. Affected systems may continue to be used in accordance with their Instructions for Use (IFU). b. Circulate this letter to all users so that they are aware of the issue. Keep this letter with the documentation of the system until Philips inspects and, where needed, corrects your system. c. Establish an emergency protocol prior to all applicable diagnostics, interventional and minimally invasive procedures to manage the situation should you experience the issue during a procedure. d. If you detect a burning odor, smoke, or fire originating from the R-cabinet in the technical room, immediately switch off the system s main power. Do not operate or use the system further. Follow your institution s emergency procedures. e. In case the affected system has been transferred to another organization, please send a copy of this letter to that organization and inform Philips about this transfer through your local Philips representative. f. Complete and return the attached response form to Philips promptly and no later than 30 days from receipt. g. Should you experience the issue reported in this letter, please report the event to Philips through your local Philips representative.
Quantity
79 units
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, India, Italy, Japan, Kenya, Korea, Republic of, Malaysia, Mexico, Nepal, Netherlands, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Yemen.
Lot, serial or model codes
Model Number: 722012; System Product Name: Allura Xper FD20; UDI-DI: 00884838059054; Serial Numbers: 2942, 1515, 1189, 1594, 535, 1964, 2156, 2742, 2092, 1673, 1803, 1655, 2020, 1973, 2037, 1853, 2098, 1248, 1082, 1576, 1057, 1129, 646, 930, 1421, 462, 203, 627, 2005, 1428, 353, 288, 191, 1994, 1802, 156, 853, 1283, 1956, 2361, 1980, 872, 931, 1295, 2513, 1090, 1374, 468, 1822, 812, 1379, 1817, 1580, 1050, 309, 144, 1841, 1845, 1468, 558, 1968, 1959, 2168, 1168, 1562, 967, 2114, 47, 167, 2091, 2053, 604, 603, 217, 2918, 373, 850, 1820, 1821, 1619;
510(k) numbers
["K102005","K130638","K130842","K133292","K161563","K141979"]
PMA numbers
not on file
Product code
OWB
Firm FEI
3002807880
Firm city
Best
Firm state
not on file
Firm country
Netherlands

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register