Skip to the content

FDA Medical Device Recalls (openFDA)

OriGen Biomedical, Inc.: Accessory Sets Syringes: 6ml syringe with a female luer with 15cm sealed SCD tubing, REF: RF-T15, and 60ml syringe on a female luer with 15cm sealed SCD tubing, REF: 15-RF60-T

Recall number
Z-1068-2024
Recalling firm
OriGen Biomedical, Inc.
Product
Accessory Sets Syringes: 6ml syringe with a female luer with 15cm sealed SCD tubing, REF: RF-T15, and 60ml syringe on a female luer with 15cm sealed SCD tubing, REF: 15-RF60-T
Status
Open, Classified
Initiated
2023-12-19
Posted
2024-02-09
Terminated
not on file
Reason
Accessory sets contain a syringes not yet qualified for the accessory's intended purpose; due to design changes to the syringe barrels and a polypropylene resin change that resulted in loss of gamma compatibility, for accessory set syringes that were gamma sterilized, which resulted in yellowed syringes.
Root cause
Nonconforming Material/Component
Action
On 11/19/23, recall notices were distributed to customers who were asked to do the following: 1) Identify and quarantine affected devices. 2) Further distribution or use should cease immediately. 3) Return affected devices to the recalling firm. 4) Complete and return the customer reply form via email to info@origen.com Customers with questions can contact the recalling firm at 1-800-233-9014 between the hours of 08:00 and 16:30 (CST). On 1/3/24, the firm sent out an additional recall communication removing one of the affected lot numbers.
Quantity
3613 Cases
Distribution
US: FL, WA, NJ, MA, OH, CA, MD, TX, NE, ND OUS: GB, CA,NL,CH, PL, DE
Lot, serial or model codes
REF/UDI-DI/Lots: RF-T15/10816203020205/ V23277, V23278, V23639; 15-RF60-T/10816203020571/V23281
510(k) numbers
not on file
PMA numbers
not on file
Product code
KJG
Firm FEI
1000160256
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register