FDA Medical Device Recalls (openFDA)
MEDLINE INDUSTRIES, LP - Northfield: Sterile Procedural Trays, labeled as the following: a. PORT/TRACH INSERTION CDS-3 b. RESIDENT DERM SURG PACK c. THROMBECTOMY PACK-LF d. AV FISTULA PACK e. AV SHUNT PACK-LF f. RR-VP SHUNT PACK-LF g. VEIN HARVESTING PACK-LF h. AV FISTULA PACK-LF i. PK, GEN-BARIATRIC j. VEIN PACK k. VENOUS ACCESS PACK-LF l. VENOUS ACCESS PACK m. VASCULAR ACCESS PACK n. ARTERIOGRAM PACK o. AV FISTULA/GRAFT INSER PK p. VEIN ABLATION PACK q. PORT PACK r. MINOR VASCULAR s. FISTULA PACK t. AV FISTULA-LF u. AV FISTULA v. PORT VENOUS ACCESS w. AV FISTULA
- Recall number
- Z-1068-2023
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Sterile Procedural Trays, labeled as the following: a. PORT/TRACH INSERTION CDS-3 b. RESIDENT DERM SURG PACK c. THROMBECTOMY PACK-LF d. AV FISTULA PACK e. AV SHUNT PACK-LF f. RR-VP SHUNT PACK-LF g. VEIN HARVESTING PACK-LF h. AV FISTULA PACK-LF i. PK, GEN-BARIATRIC j. VEIN PACK k. VENOUS ACCESS PACK-LF l. VENOUS ACCESS PACK m. VASCULAR ACCESS PACK n. ARTERIOGRAM PACK o. AV FISTULA/GRAFT INSER PK p. VEIN ABLATION PACK q. PORT PACK r. MINOR VASCULAR s. FISTULA PACK t. AV FISTULA-LF u. AV FISTULA v. PORT VENOUS ACCESS w. AV FISTULA
- Status
- Open, Classified
- Initiated
- 2022-11-17
- Posted
- 2023-02-06
- Terminated
- not on file
- Reason
- Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Root cause
- Device Design
- Action
- Medline Industries, LP notified consignees via email and first-class mail on 11/15/2022. Consignees were instructed to check stock for affected product and immediately quarantine. For any affected units, customers were instructed to over-label the sterile kits with a warning label indicating "Warning To Avoid Injury, Please Handle Blades with Care." No product is to be returned. Customers were also instructed to complete and return the response form and notify customers if further distributed.
- Quantity
- 622 cases (1,619 units)
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Japan, Mexico, Panama, Turks & Caicos, United Arab Emirates.
- Lot, serial or model codes
- a. PORT/TRACH INSERTION CDS-3, Item Number: CDS983004D, Case UDI/GTIN: 40889942000464, Unit UDI/GTIN: 10889942000463, Lot Number: 22IBF922,22IBE764; b. RESIDENT DERM SURG PACK, Item Number: DYNDA2997, Case UDI/GTIN: 40195327235353, Unit UDI/GTIN: 10195327235352, Lot Number: 22JBR611; c. THROMBECTOMY PACK-LF, Item Number: DYNJ0160348B, Case UDI/GTIN: 40889942211594, Unit UDI/GTIN: 10889942211593, Lot Number: 22IDB621; d. AV FISTULA PACK, Item Number: DYNJ01966N, Case UDI/GTIN: 40193489383769, Unit UDI/GTIN: 10193489383768, Lot Number: 22IMA429; e. AV SHUNT PACK-LF, Item Number: DYNJ0781844W, Case UDI/GTIN: 40193489226530, Unit UDI/GTIN: 10193489226539, Lot Number: 22IMG652; f. RR-VP SHUNT PACK-LF, Item Number: DYNJ0785917M, Case UDI/GTIN: 40193489412827, Unit UDI/GTIN: 10193489412826, Lot Number: 22GMH382; g. VEIN HARVESTING PACK-LF, Item Number: DYNJ0826732I, Case UDI/GTIN: 40193489377973, Unit UDI/GTIN: 10193489377972, Lot Number: 22IMF839; h. AV FISTULA PACK-LF, Item Number: DYNJ0826753O, Case UDI/GTIN: 40193489384292, Unit UDI/GTIN: 10193489384291, Lot Number: 22JMH548,22IMC908; i. PK, GEN-BARIATRIC, Item Number: DYNJ33890B, Case UDI/GTIN: 40193489213455, Unit UDI/GTIN: 10193489213454, Lot Number: 22JBV659,22HBW898; j. VEIN PACK, Item Number: DYNJ36309F, Case UDI/GTIN: 40195327078035, Unit UDI/GTIN: 10195327078034, Lot Number: 22IMC227; k. VENOUS ACCESS PACK-LF, Item Number: DYNJ36725D, Case UDI/GTIN: 40195327148714, Unit UDI/GTIN: 10195327148713, Lot Number: 22HBO043; l. VENOUS ACCESS PACK, Item Number: DYNJ42694A, Case UDI/GTIN: 40193489983341, Unit UDI/GTIN: 10193489983340, Lot Number: 22JBK167,22IBJ914; m. VASCULAR ACCESS PACK, Item Number: DYNJ45071C, Case UDI/GTIN: 40889942943303, Unit UDI/GTIN: 10889942943302, Lot Number: 22IBO914; n. ARTERIOGRAM PACK, Item Number: DYNJ53095A, Case UDI/GTIN: 40195327148721, Unit UDI/GTIN: 10195327148720, Lot Number: 22IBM491,22HBY446; o. AV FISTULA/GRAFT INSER PK, Item Number: DYNJ57152, Case UDI/GTIN: 40889942721888, Unit UDI/GTIN: 10889942721887, Lot Number: 22IBT472; p. VEIN ABLATION PACK, Item Number: DYNJ62889B, Case UDI/GTIN: 40195327030958, Unit UDI/GTIN: 10195327030957, Lot Number: 22JDB544; q. PORT PACK, Item Number: DYNJ64675, Case UDI/GTIN: 40193489212144, Unit UDI/GTIN: 10193489212143, Lot Number: 22IBP112; r. MINOR VASCULAR, Item Number: DYNJ66148A, Case UDI/GTIN: 40195327096329, Unit UDI/GTIN: 10195327096328, Lot Number: 22JBL900,22HBP630,22HBG264; s. FISTULA PACK, Item Number: DYNJ66502, Case UDI/GTIN: 40193489358507, Unit UDI/GTIN: 10193489358506, Lot Number: 22HBO053; t. AV FISTULA-LF, Item Number: DYNJ901747K, Case UDI/GTIN: 40193489482165, Unit UDI/GTIN: 10193489482164, Lot Number: 22JDC010,22JDA876; u. AV FISTULA, Item Number: DYNJ903417F, Case UDI/GTIN: 40193489418218, Unit UDI/GTIN: 10193489418217, Lot Number: 22KBF261,22KBG353; v. PORT VENOUS ACCESS, Item Number: DYNJ903426F, Case UDI/GTIN: 40193489418447, Unit UDI/GTIN: 10193489418446, Lot Number: 22JBR462,22JBB174,22HBS253; w. AV FISTULA, Item Number: DYNJ903726B, Case UDI/GTIN: 40193489966849, Unit UDI/GTIN: 10193489966848, Lot Number: 22KBD044,22IBO370,22HBX959;
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OFF
- Firm FEI
- 1417592
- Firm city
- Northfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28