Natus Medical Incorporated: EPWorks software used in the Xltek Protektor Stimulator Product Usage: Intraoperative diagnosis of acute dysfunction in corticospinal axonal conduction brought about by mechanical trauma (traction, shearing, laceration, or compression) or vascular insufficiency.
- Recall number
- Z-1068-2015
- Recalling firm
- Natus Medical Incorporated
- Product
- EPWorks software used in the Xltek Protektor Stimulator Product Usage: Intraoperative diagnosis of acute dysfunction in corticospinal axonal conduction brought about by mechanical trauma (traction, shearing, laceration, or compression) or vascular insufficiency.
- Status
- Terminated
- Initiated
- 2015-01-02
- Posted
- 2015-02-09
- Terminated
- 2017-03-22
- Reason
- Software error occurs when using remote monitoring; if the remote user tries to stop the free run waveform group, the system will display a message informing the user that they do not have sufficient privilege.
- Root cause
- Software design
- Action
- Natus Medical Inc. DBA Excel-Tech Ltd (Xltek) issued medical device notification letters dated December 22, 2014 to their customers vial email and via courier service on January 2, 2015. Within the notice, customers are instructed are on to removal the previous installation and also on how to install the new program upgrade. Natus Medical Inc. has requested that the Previous Version Removed from Use Confirmation form be completed by each customer and returned. Customers who require assistance can contact Natus Technical service by calling 800-387-7516 or via email Oakville_Technical_Service@xltek.com or ots@natus.com.
- Quantity
- 928 units
- Distribution
- US Nationwide Distribution
- Lot, serial or model codes
- EPWorks Software Versions prior to 6.0
- 510(k) numbers
- ["K040358"]
- PMA numbers
- not on file
- Product code
- GWF
- Firm FEI
- 3002809369
- Firm city
- Oakville Canada
- Firm state
- Ontario
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28