FDA Medical Device Recalls (openFDA)
Smith & Nephew Inc.: Renasys EZ PLUS, a Canister Component of of Renasys EZ Negative Pressure Wound Therapy System
- Recall number
- Z-1068-2014
- Recalling firm
- Smith & Nephew Inc.
- Product
- Renasys EZ PLUS, a Canister Component of of Renasys EZ Negative Pressure Wound Therapy System
- Status
- Terminated
- Initiated
- 2013-11-14
- Posted
- 2014-02-21
- Terminated
- 2014-08-04
- Reason
- Modification of the bacterial overflow guard (filter) and related changes to product labeling.
- Root cause
- Device Design
- Action
- The correction strategy is as follows: " An Urgent Medical Device Correction Notice has been prepared to be distributed to all US consignees via Federal Express delivery for confirmation of the delivery. " A copy of the Urgent Medical Device Correction Notice is to be posted on Smith & Nephew's websites, URL corporate, http://www.smith-nephew.com/news-and-media/, and myrenasys.com, concurrent with the distribution of the Notice to consignees. " US distribution of the original design of Canister Kits codes (66800912, 66800913, 66801066, 66800423 and 66800058) subject to this correction ceased November 1, 2013. " The correction has been implemented on a change being affected basis. The modified design is being distributed to US customers effective November 1, 2013. No product will be removed from the market. " Periodic progress reports to FDA, as required per 21 CFR 806 will be submitted monthly until completion of the correction actions..
- Quantity
- 6,871,375 canisters
- Distribution
- Worldwide Distribution: USA (nationwide) including states of: AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KY, LA, MA, MI, MN, MO, MS, MT, NC, NH, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, VA, WA, WV, and Puerto Rico; and the countries of: Austria, Australia, Belgium, Switzerland, China, Cyprus, Germany, Denmark, Spain, Finland, France, UK, Greece, Israel, Italy, Malta, Netherlands, Norway, Paraguay, Russia, Saudi Arabia, Sweden, Thailand, Turkey, South Africa, Mexico, and Canada.
- Lot, serial or model codes
- Product code: 66800912 - 800ml canister with solidifier 66800913 - 250ml canister with solidifier 66801066 - 800ml canister without solidifier 66800423 - 800ml canister with solidifier 66800058 - 250ml canister with solidifier Lot codes - M201498 thru M400299
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OMP
- Firm FEI
- 3006760724
- Firm city
- Saint Petersburg
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28