Skip to the content

FDA Medical Device Recalls (openFDA)

Ethicon Endo-Surgery Inc: ENDOPATH ETS-FLEX45 Articulating Endoscopic Linear Cutters, Product Codes: ATS45; LONG45A; 6TB45; ATG45; ATB45; ATW45; ATS45NK; and NAW45. Intended for transection, resection, and/or creation of anastomoses.

Recall number
Z-1068-2011
Recalling firm
Ethicon Endo-Surgery Inc
Product
ENDOPATH ETS-FLEX45 Articulating Endoscopic Linear Cutters, Product Codes: ATS45; LONG45A; 6TB45; ATG45; ATB45; ATW45; ATS45NK; and NAW45. Intended for transection, resection, and/or creation of anastomoses.
Status
Terminated
Initiated
2009-03-27
Posted
2011-02-15
Terminated
2013-02-15
Reason
The firm initiated this second subsequent recall because the affected devices can remain closed and clamped on tissue during surgical procedures due to the fact that the mechanism which connects the articulation joint to the device shaft is insufficient.
Root cause
Other
Action
Ethicon Endo-Surgery issued "Urgent Expanded Device Recall Event 1981" notification letters dated April 7, 2009, to their customers. They provided a list of all Affected Product Codes of the ETS FLEX45 Articulating Endoscopic Linear Cutters (including the Product Code: ATS45 from the 1st recall) and Affected Flex Tray Codes subject to recall. The notification document reminds the Operating Room and Materials Management customers of the problem which necessitated the previous recall (Z-0361-2011). It goes on to instruct their customers to immediately discontinue use of the affected products. They are to complete and fax back the Business Reply Form, Affected product is to be returned with a copy of the completed Business Reply Form to Stericycle, Replacement product should be received within 30-60 days after return, Customers can contact Stericycle concerning this recall at 1-888-365-4680, reference Event 1981.
Quantity
29,403 units
Distribution
Nationwide Distribution -- USA including the following states, AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MO, MS, MT, NC, NE, NH, NJ, MN, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, and Puerto Rico.
Lot, serial or model codes
The recalled Lot Numbers are: E4MJ47, E4MF6W, E4MK84, E4MM23, E4MM64, E4MP4A, E4MR15, E4MT07, E4MT4Y, F4MU2R, F4MY91, E4MG53, E4MG5K, E4MG5L, E4MN39, E4MG89, E4MG9M, E4MH2L, E4ML6J, E4MG77, E4MG8A, E4MH18, E4MJ3K, F4MX8C, and F4N799.
510(k) numbers
["K020779","K070887"]
PMA numbers
not on file
Product code
GEI
Firm FEI
1527736
Firm city
Cincinnati
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register