Skip to the content

FDA Medical Device Recalls (openFDA)

Toshiba American Medical Systems Inc: TSX-IOI: AQUILION 32/64; System, X-Ray, Tomography, Computed. Designed to produce cross-sectional images of a human body.

Recall number
Z-1067-2013
Recalling firm
Toshiba American Medical Systems Inc
Product
TSX-IOI: AQUILION 32/64; System, X-Ray, Tomography, Computed. Designed to produce cross-sectional images of a human body.
Status
Terminated
Initiated
2011-11-30
Posted
2013-04-08
Terminated
2013-06-17
Reason
The firm initiated this recall due to a potential software issue. The DLP (Dose Length Product) value may be displayed incorrectly when the vHP option is used.
Root cause
Process design
Action
Toshiba sent "URGENT: MEDICAL DEVICE CORRECTION" letters to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. The letter contained a return reply form that is to be faxed to TAMS for retention. Contact the firm at (800) 421-1968 for questions relating to this letter.
Quantity
23 units
Distribution
Nationwide Distribution including the states of WA, OH, CA, TX, WV, IL, AZ, CA, WI, ME, WI, NC, FL, MT, and HI.
Lot, serial or model codes
Serial Numbers: HDA0722595 JDA0972130 JDA0972135 JDA0992147 JDA09Y2158 KDA09X2063 JDA09Y2157 JDA09Y2161 JDA1012171 JDA10X2191 JDA10X2197 JDA10Z2214 JGAl112220 JLA1122232 JLA1132235 JDAl112216 JDAl112219 NLA0972032 JDA1182242 JDA1182241 JDA1182244 JDA1072183 JDA1082189
510(k) numbers
not on file
PMA numbers
not on file
Product code
JAK
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register