FDA Medical Device Recalls (openFDA)
MEDLINE INDUSTRIES, LP - Northfield: Sterile Procedural Trays, labeled as the following: a. CARDIAC CATH PACK-LF; b. ANGIOGRAPHY-NS-LF; c. EP BASIC SET UP PACK; d. IMPLANT PACK-LF; e. PORT PACEMAKER INSERTION PACK; f. OPEN HEART PACK; g. LAP BASIN TRAY; h. ANGIO ACCESS; i. ANGIO
- Recall number
- Z-1066-2023
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Sterile Procedural Trays, labeled as the following: a. CARDIAC CATH PACK-LF; b. ANGIOGRAPHY-NS-LF; c. EP BASIC SET UP PACK; d. IMPLANT PACK-LF; e. PORT PACEMAKER INSERTION PACK; f. OPEN HEART PACK; g. LAP BASIN TRAY; h. ANGIO ACCESS; i. ANGIO
- Status
- Open, Classified
- Initiated
- 2022-11-17
- Posted
- 2023-02-06
- Terminated
- not on file
- Reason
- Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Root cause
- Device Design
- Action
- Medline Industries, LP notified consignees via email and first-class mail on 11/15/2022. Consignees were instructed to check stock for affected product and immediately quarantine. For any affected units, customers were instructed to over-label the sterile kits with a warning label indicating "Warning To Avoid Injury, Please Handle Blades with Care." No product is to be returned. Customers were also instructed to complete and return the response form and notify customers if further distributed.
- Quantity
- 314 cases (719 units)
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Japan, Mexico, Panama, Turks & Caicos, United Arab Emirates.
- Lot, serial or model codes
- a. CARDIAC CATH PACK-LF, Item Number: DYNJ0368386D, Case UDI/GTIN: 40889942060246, Unit UDI/GTIN: 10889942060245, Lot Number: 22IMD669; b. ANGIOGRAPHY-NS-LF, Item Number: DYNJ24913N, Case UDI/GTIN: 40195327127368, Unit UDI/GTIN: 10195327127367, Lot Number: 22JBC783; c. EP BASIC SET UP PACK, Item Number: DYNJ27920D, Case UDI/GTIN: 40884389855349, Unit UDI/GTIN: 10884389855348, Lot Number: 22IBS859; d. IMPLANT PACK-LF, Item Number: DYNJ43426C, Case UDI/GTIN: 40195327255313, Unit UDI/GTIN: 10195327255312, Lot Number: 22IBO525; e. PORT PACEMAKER INSERTION PACK, Item Number: DYNJ43506A, Case UDI/GTIN: 40888277541000, Unit UDI/GTIN: 10888277541009, Lot Number: 22IBT230; f. OPEN HEART PACK, Item Number: DYNJ43937, Case UDI/GTIN: 40888277077967, Unit UDI/GTIN: 10888277077966, Lot Number: 22HBR302; g. LAP BASIN TRAY, Item Number: DYNJ50908B, Case UDI/GTIN: 40193489997621, Unit UDI/GTIN: 10193489997620, Lot Number: 22JDA804; h. ANGIO ACCESS, Item Number: DYNJ82082, Case UDI/GTIN: 40195327167616, Unit UDI/GTIN: 10195327167615, Lot Number: 22JMA564; i. ANGIO, Item Number: DYNJ905151B, Case UDI/GTIN: 40195327219964, Unit UDI/GTIN: 10195327219963, Lot Number: 22JMG564,22IMG399;
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OES
- Firm FEI
- 1417592
- Firm city
- Northfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as