Skip to the content

FDA Medical Device Recalls (openFDA)

Elekta, Inc.: Elekta Synergy XVI 4.5 and R4.6 To be used as part of radiation therapy treatment process.

Recall number
Z-1066-2012
Recalling firm
Elekta, Inc.
Product
Elekta Synergy XVI 4.5 and R4.6 To be used as part of radiation therapy treatment process.
Status
Terminated
Initiated
2012-01-26
Posted
2012-02-21
Terminated
2012-04-17
Reason
If a user, using an on-line correction protocol, clicks the accept correction button, an approval dialog box appears. If cancel is selected, the table correction values return to their initial values, but the image panes continue to display the changed position.
Root cause
Software design
Action
ELEKTA sent an Important Field Safety Notice (20001507042) dated January 2012 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. For questions regarding this recall call 770-670-2548
Quantity
96 units
Distribution
Nationwide Distribution including AR, AZ, CA, CT, FL, GA, ID, IL, LA MD, MA, MI, MN, MO, NJ, NY, NC, ND, OH, OK, OR, PA, RI, TN, VA, WA, WI, and Puerto Rico
Lot, serial or model codes
152433, 152019, 151176, 152248, 152262, 152454, 105941, 151802, 152427, 151628, 151362, 152207, 151545, 151566, 152271, 152425, 152452, 152491, 151808, 152163, 152331, 151443, 151955, 152585, 152522, 151514, 151791, 151430, 151931. 105984, 151412, 152388, 151616, 151141, 152303, 151892, 151785, 152358, 152223, 152305, 152468, 151260, 152469, 152481, 152601, 152356, 152175, 152365, 152177, 152257, 151952, 152353, 152359, 152362, 152352, 152371, 152370, 151942, 151820, 151770, 151864, 151396, 152448, 151954, 151076, 151309, 152103, 152301, 151634, 151793, 152210, 152211, 151080, 152373, 152222, 152158, 152392, 151684, 152530, 152536, 151168, 151839, 151156, 151155, 151157, 152117, 151710, 152308, 152307, 152504, 151683, 151764, 151885, 152432, 151812, 152549.
510(k) numbers
["K100115"]
PMA numbers
not on file
Product code
IYE
Firm FEI
1037831
Firm city
Norcross
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register