Skip to the content

FDA Medical Device Recalls (openFDA)

SCC Soft Computer: SoftPath Laboratory Information System. Versions 4.3.0.8, 4.3.0.9, 4.3.0.10, 4.3.0.11, 4.3.0.12, 4.3.0.14, 4.3.0.15, and 4.4.0.0

Recall number
Z-1065-2015
Recalling firm
SCC Soft Computer
Product
SoftPath Laboratory Information System. Versions 4.3.0.8, 4.3.0.9, 4.3.0.10, 4.3.0.11, 4.3.0.12, 4.3.0.14, 4.3.0.15, and 4.4.0.0
Status
Terminated
Initiated
2014-11-26
Posted
2015-02-09
Terminated
2016-05-16
Reason
Modifications to diagnostic text may be: 1) Saved to the database but not appear on the report sent to the physician; or 2) Documented on the report, but not saved to the database.
Root cause
Software design
Action
On 11/26/2014 SCC Soft Computer sent a Correction Communication to their consignees: Corrective Action: Versions 4.3.0.8 through 4.4.0.2 will have a mandatory hotfix created. Due to the amount of versions affected, the range of availability dates will be 11/24/2014 through 12/30/2014. Utility: A utility will be made available to be run on your Live system that will identify cases in which there is a disparity between the diagnosis text in the document and the database. The utility output will be provided for client review and validation.
Quantity
108 units
Distribution
Distribution US Nationwide and CANADA.
Lot, serial or model codes
Version 4.3.0.8 Release date: 07/31/2009 - Present;  Version 4.3.0.9 Release date: 11/23/2009 - Present;  Version 4.3.0.10 Release date: 09/20/2010- Present;  Version 4.3.0.11 Release date: 11/11/2010- Present;  Version 4.3.0.12 Release date: 02/07/2011- Present; Version 4.3.0.14 Release date: 11/14/2012- Present;  Version 4.3.0.15 Release date: 09/17/2013- Present;  Version 4.4.0.1 Release date: 12/13/2013 - Present;
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQP
Firm FEI
1000307651
Firm city
Clearwater
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register