Soleil Systems, Inc.: Sollux Systeme commercial tanning beds; Soleil Systems, Inc. The following models are affected by this recall: a) Sollux Systeme 32 - 32 x 100 w tanning bulbs, no facial tanner bulbs b) Sollux Systeme 32.3 - 32 x 100 w tanning bulbs, plus 3 x 400 w facial tanner bulbs c) Sollux Systeme 36 - 18 x 100 w reflective tanning bulbs, 18 x 100 w tanning bulbs, no facial tanner bulbs d) Sollux Systeme 36.3 - 18 x 100 w reflective tanning bulbs, 18 x 100 w tanning bulbs, plus 3 x 400 w facial tanner bulbs e) Sollux System 40 - 20 x 160 w reflective tanning bulbs, 20 x 100 w reflective tanning bulbs, no facial tanner bulbs f) Sollux System 40.3 - 20 x 160 w reflective tanning bulbs, 20 x 100 w reflective tanning bulbs, 3 x 400 w facial tanner bulbs

Recall number
Z-1065-04
Recalling firm
Soleil Systems, Inc.
Product
Sollux Systeme commercial tanning beds; Soleil Systems, Inc. The following models are affected by this recall: a) Sollux Systeme 32 - 32 x 100 w tanning bulbs, no facial tanner bulbs b) Sollux Systeme 32.3 - 32 x 100 w tanning bulbs, plus 3 x 400 w facial tanner bulbs c) Sollux Systeme 36 - 18 x 100 w reflective tanning bulbs, 18 x 100 w tanning bulbs, no facial tanner bulbs d) Sollux Systeme 36.3 - 18 x 100 w reflective tanning bulbs, 18 x 100 w tanning bulbs, plus 3 x 400 w facial tanner bulbs e) Sollux System 40 - 20 x 160 w reflective tanning bulbs, 20 x 100 w reflective tanning bulbs, no facial tanner bulbs f) Sollux System 40.3 - 20 x 160 w reflective tanning bulbs, 20 x 100 w reflective tanning bulbs, 3 x 400 w facial tanner bulbs
Status
Terminated
Initiated
2004-08-02
Posted
not on file
Terminated
2006-03-28
Reason
The sunlamp products failed to incorporate a timer system and a control for termination of radiation emission in accordance with the Federal Sunlamp Product Performance Standard.
Root cause
Other
Action
Soleil Systems sent recall notification letters dated 6/28/04 to their dealers on 8/2/04, informing them of the noncompliances of failing to incorporate a permanently mounted timer and on/off switch in the sunbeds. Soleil will bring each sunbed into compliance and requested the dealers to review their sales records for any Sollux beds received between 4/28/04 and 11/01/03, and supply Soleil with a list of names and addresses of end users who received those units. Soleil sent out end user notifications dated 8/20/04 beginning on 8/23/04 and initiated the retrofit of the sunbeds on 10/25/04.
Quantity
70 beds
Distribution
The sunbeds were distributed through dealers in Missouri, Texas, Virginia, Kentucky and Illinois
Lot, serial or model codes
Each sunbed has two serial numbers, one on the canopy and another on the bench. The serial numbers listed below list the canopy/bench.serial numbers followed by the manufacturing date.  a) Sollux Systeme 32 - 22 units, serial numbers 320002/320007 May 03; 320008/320014 May 03; 320010/320010 May 03; 320012/320012 May 03; 320015/321207 May 03; 320016/320016 May 03; 320017/320017 May 03; 320018/320018 May 03; 320019/320019 May 03; 320020/321202 May 03; 320021/321211 May 03; 320095/321264 Oct 03; 320096/321325 Oct 03; 320104/321284 Jun 03; 320106/321320 Jun 03; 320109/320109 May 03; 320111/321319 Aug 03; 320112/321314 Aug 03; 320123/321317 Sep 03; 320214/320214 May 03; 321211/321211 May 03; 321324/320113 Aug 03;   b) Sollux Systeme 32.3 - 7 units, serial numbers 320007/321104 May 03; 320008/321109 May 03; 320009/321112 May 03; 320011/321316 Aug 03; 320093/321269 Oct 03; 320126/321268 Oct 03; 320127/321323 Oct 03;   c) Sollux Systeme 36 - 8 units, serial numbers  360004/360016 Sep 03; 360009/360015 May 03; 360012/361003 May 03; 360013/360021 Jun 03; 361001/360013 Aug 03; 361003/360017 Aug 03; 361041/361004 Oct 03; 361042/361001 Aug 03;   d) Sollux Systeme 36.3 - 8 units, serial numbers 360011/360019 May 03; 360012/360015 May 03; 360013/360020 Jun 03; 360015/360010 May 03; 360016/360011 May 03; 360018/360014 My 03; 360020/360023 May 03; 360021/360027 May 03;  e) Sollux System 40 - 1 unit, serial number  400002/400008 May 03;   f) Sollux System 40.3 - 24 units, serial numbers 400002/400010 May 03; 400003/400021 Jul 03; 400004/400005 May 03; 400005/400020 Jun 03; 400006/400014 Jun 03; 400007/400004 Jun 03; 400008/400019 Aug 03; 400009/400094 Aug 03; 400010/400005 Jul 03; 400011/400023 Aug 03; 400012/400016 Apr 03; 400013/400017 Oct 03; 400116/400001 Sep 03; 400019/400020 Oct 03; 400021/400013 Aug 03; 400090/400015 Aug 03; 400095/400010 Aug 03; 400096/400021 Sep 03; 400097/400013 Oct 03; 400099/400018 Sep 03; 400102/400015 Sep 03; 400113/400012 Jul 03; 400114/400002 JUL 03; 400115/400009 Jul 03;
510(k) numbers
not on file
PMA numbers
not on file
Product code
U--19
Firm FEI
3004002126
Firm city
Elk Grove Village
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register