Skip to the content

FDA Medical Device Recalls (openFDA)

Keystone Dental Inc: Paltop 1.25 Hex Drivers, Short Part Number: 60-70102

Recall number
Z-1064-2019
Recalling firm
Keystone Dental Inc
Product
Paltop 1.25 Hex Drivers, Short Part Number: 60-70102
Status
Terminated
Initiated
2019-02-22
Posted
not on file
Terminated
2020-06-19
Reason
Paltop Surgical Kits contain 1.25mm Hex Driver(s) which may have a slightly oversized hex driving feature, make it difficult to engage screws and abutments containing the corresponding screw feature
Root cause
Process control
Action
Keystone Dental initiated recall by phone to each one of the impacted customers. The phone call was then followed by a Recall Notification Letter that was mailed via overnight FedEx mail on February 22, 2019. Letters states problem and action to take: Replacement 1.25mm Hex Driver(s) have been provided to you with this letter. The 1.25mm Hex Driver(s) currently in your kit should be discarded. Please refer to the images on page 2 to locate the driver within your kit. This issue is limited to this 1.25mm Hex Driver(s) only and no other product within the kit is affected by this issue. Complete the Reply Communication form.
Quantity
3 Hex Driver
Distribution
FL, MA, SC, VA
Lot, serial or model codes
Lot Number: WO-010354
510(k) numbers
["K112795"]
PMA numbers
not on file
Product code
DZE
Firm FEI
3005990499
Firm city
Burlington
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register