Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Siemens syngo Imaging XS. Radiological image processing system.

Recall number
Z-1064-2012
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Siemens syngo Imaging XS. Radiological image processing system.
Status
Terminated
Initiated
2012-01-26
Posted
2012-02-20
Terminated
2012-09-29
Reason
In September 2011 firm issued two Customer Safety Advisory Notices for syngo Imaging XS. Firm became aware of an unintended behavior when using syngo Imaging XS, version VA70A and VA70B with Client/Server Installations. Loading additional images/series from the Patient Browser using the thumbnails may cause unintended deletion of images/series. In response, firm issued a Customer Safety Advisor
Root cause
Software design
Action
Siemens Medical issued an "UPDATE INSTRUCTIONS SY055/11/S/UPDATE PACKAGE VA70B_0211" to all affected customers. The package described the product, problem, and instructions for customers to follow.
Quantity
128
Distribution
Nationwide Distribution-including the states of CA, CO, CT, FL, GA, IL, IN, IA, KS, MD, MN, MO, NE, NV, NJ, NY, NC, OH, PA, TN, TX, and WI.
Lot, serial or model codes
Model number 10496279 -- serial numbers 1344, 1744, 1187, 2981, 1361, 3052, 3053, 3054, 3055, 3056, 3057, 3058, 3059, 3060, 3061, 2323, 2254, 1088, 1089, 2194, 2195, 1213, 1214, 1188, 1189, 3086, 2626, 1828, 1624, 1625, 2878, 2365, 2366, 3087, 3088, 3089, 2500, 2971, 1424, 1425, 3093, 2890, 1371, 1372, 3062, 2267, 2268, 2269, 1856, 1857, 2695, 2696, 3010, 1373, 1374, 1375, 1376, 1377, 1713, 1712, 2408, 2409, 2559, 2560, 2561, 2562, 1735, 1668, 2098, 1190, 1418, 2069, 2698, 2769, 1884, 1381, 1382, 1378, 1379, 1380, 1165, 2310, 2311, 2312, 2448, 2449, 2450, 2670, 2671, 2672, 2673, 2674, 2675, 2676, 2677, 2678, 2679, 2324, 1940, 1395, 2901, 2897, 2898, 2899, 2900, 2934, 3017, 2144, 2145, 2146, 2147, 2148, 2149, 2150, 2151, 2152, 2153, 2154, 2155, 2156, 2157, 2158, 2159, 2200, 2201, 1383, 3050, and 3085.
510(k) numbers
["K082430"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register