FDA Medical Device Recalls (openFDA)
Zimmer Inc.: Zimmer NexGen Complete Knee Solution Minimally Invasive Solutions Procedures Trabecular Metal Technology Tibial Tray, fixed bearing, size 8, sterile, Zimmer, Warsaw, Indiana; REF 00-5954-057-02. The product is used as an implant for knee replacement.
- Recall number
- Z-1064-2010
- Recalling firm
- Zimmer Inc.
- Product
- Zimmer NexGen Complete Knee Solution Minimally Invasive Solutions Procedures Trabecular Metal Technology Tibial Tray, fixed bearing, size 8, sterile, Zimmer, Warsaw, Indiana; REF 00-5954-057-02. The product is used as an implant for knee replacement.
- Status
- Terminated
- Initiated
- 2009-11-16
- Posted
- 2010-03-10
- Terminated
- 2011-03-23
- Reason
- The titanium portion of the implant may separate from the trabecular metal material.
- Root cause
- Other
- Action
- On December 1, 2009, an e-mail was sent to consignees for lot 60866613 to quarantine and return the product. On December 2, 2009 an e-mail was sent to U.S. distributors, and on December 3, 2009 and e-mail was sent to international distributors, instructing them to suspend usage and return all lots of size 8 trays. On December 9, 2009, a recall letter was sent to U.S. user accounts who received the products directly from Zimmer, U.S. distributors and Zimmer's U.S. sales force informing them of the recall. Direct questions about this recall to Zimmer, Inc. by calling 1-800-613-6131.
- Quantity
- 846
- Distribution
- Worldwide Distribution -- United States, Australia, Austria, Belgium, Canada, Denmark, France, Germany, Italy, Saudi Arabia, Spain, Sweden, Switzerland and the United Kingdom.
- Lot, serial or model codes
- All lots. Lots 00111134, 60747593, 60747594, 60752676, 60752677, 60760248, 60768076, 60821507, 60824004, 60830411, 60830412, 60845351, 60850604, 60852670, 60852671, 60860895, 60866613, 60872311, 60879050, 60892605, 60900745, 60907449, 60917234, 60922494, 60938889, 60938890, 60947027, 60947028, 60963035, 60963036, 60963037, 60963039, 60983254, 61015851, 61028168, 61033723, 61033724, 61068387, 61068388, 61068389, 61074754, 61074755, 61084771, 61096604, 61116246, 61116247, 61137056, 61143427, 61149041, 61157940, 61173295, 61193516, 61193520, 61198179, 61204038, 61215644, 61225562, 61235612, 61240323, 61248321, 61259184, 61263170, 61272449, 61285508, 61298201, 61306497, 61280371, 61315873, 61319581 and 61329266.
- 510(k) numbers
- ["K072160"]
- PMA numbers
- not on file
- Product code
- JWH
- Firm FEI
- 1000220733
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28