Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Laboratories: CELL-Dyn Ruby Hematology Analyzer, List Number 08H67-01 The CELL-DYN Ruby is a multi-parameter, automated hematology analyzer designed for in vitro diagnostic use in clinical laboratories. The instrument has two modes of operation: the Autoloader mode that aspirates samples from closed collection tubes and the Open Tube mode that aspirates samples from open collection tubes.

Recall number
Z-1064-2009
Recalling firm
Abbott Laboratories
Product
CELL-Dyn Ruby Hematology Analyzer, List Number 08H67-01 The CELL-DYN Ruby is a multi-parameter, automated hematology analyzer designed for in vitro diagnostic use in clinical laboratories. The instrument has two modes of operation: the Autoloader mode that aspirates samples from closed collection tubes and the Open Tube mode that aspirates samples from open collection tubes.
Status
Terminated
Initiated
2008-11-17
Posted
2009-04-10
Terminated
2010-11-27
Reason
Carryover failures in software revisions 1.0ML and 2.0ML result in elevated PLT background count.
Root cause
Other
Action
Recall initiated on 11/17/2008. A Product Correction letter and customer reply form is sent to all currently active CELL-DYN Ruby customers. The customers are asked to keep the communication with their CELL-DYN Ruby System Operator's Manual and to note that the information listed in the table replaces the carryover specification provided in all applications of the manual. Questions from customers in the US should be directed to Customer Support at 1-877-4ABBOTT. Customers outside the US should contact their local hematology customer support representative.
Quantity
899 units
Distribution
Worldwide Distribution
Lot, serial or model codes
Serial No. 34001BG - 34965BG
510(k) numbers
["K061667"]
PMA numbers
not on file
Product code
GKZ
Firm FEI
2919069
Firm city
Santa Clara
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register