FDA Medical Device Recalls (openFDA)
Villa Radiology Systems LLC: Juno DRF; Model: 709020 Multi-function x-ray systems capable of routine radiograph and general fluoroscopy procedures.
- Recall number
- Z-1063-2017
- Recalling firm
- Villa Radiology Systems LLC
- Product
- Juno DRF; Model: 709020 Multi-function x-ray systems capable of routine radiograph and general fluoroscopy procedures.
- Status
- Open, Classified
- Initiated
- 2016-02-19
- Posted
- not on file
- Terminated
- not on file
- Reason
- The analysis showed that the incident was caused by a chain of four faults: 1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position. 2) The directions for use (device labeling and instructions for use) did not contain an explicit exclusion of the prohibited area. 3.) The movement was activated inadvertently and continuously. and it was not released (the movement requires continuous activation). 4.) Delayed activation of the emergency stop.
- Root cause
- Device Design
- Action
- Consignees were notified of the recall via letter starting on 2/19/16. The letter explains the issue and actions to be taken by the consignee and manufacturer.
- Quantity
- 66 distributed in US
- Distribution
- AL AZ CA CO CT DC FL IA IL IN LA MA MD MI MN MO MS NJ NY OH PA Puerto Rico TN TX UT WI
- Lot, serial or model codes
- 10110770 11030805 11030806 11070845 11070848 11070849 11090864 11110900 12020978 12020982 12030999 12041003 12061014 12061018 12061020 12081040 12091074 12101104 12121186 12121187 13031215 13101263 14021301 14031313 14041321 14051332 14071353 14081359 14091368 14091371 14121400 14121403 15021426 15031440 15041444 15091473 15091478 15101484 15101487
- 510(k) numbers
- ["K050190"]
- PMA numbers
- not on file
- Product code
- KXJ
- Firm FEI
- 2243757
- Firm city
- Oxford
- Firm state
- CT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28