FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc.: ACUSON S2000 Automated Breast Volume Scanner (ABVS), model number 10434730. The affected sub-components are #10434968 - Boom Welding and 10434969 - Arm Welding. The S2000 ABVS ultrasound System for imaging the breast allows volumetric acquisition of the breast image.
- Recall number
- Z-1063-2015
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- ACUSON S2000 Automated Breast Volume Scanner (ABVS), model number 10434730. The affected sub-components are #10434968 - Boom Welding and 10434969 - Arm Welding. The S2000 ABVS ultrasound System for imaging the breast allows volumetric acquisition of the breast image.
- Status
- Terminated
- Initiated
- 2015-01-12
- Posted
- 2015-02-07
- Terminated
- 2015-06-10
- Reason
- There is a low probability the scanner arm will become completely detached from the scanner column.
- Root cause
- Vendor change control
- Action
- Siemens sent a Customer Safety Advisory Notification to all affected sites on January 12, 2015, certified mail, with return receipt. The letter identified the affected product, the issue, and potential risk. Customers were asked to immediately discontinue use of the affected product. A Customer Service Engineer will contact customers to schedule an appointment to make the necessary repairs. Questions should be directed to a local service support person for information regarding timelines and status. Customers are to share the information with all personnel, who need to be aware of the issue.
- Quantity
- 24 units
- Distribution
- Worldwide Distribution -- US, including the states of TX and MN, and the territory of P.R.; and, the countries of China, Germany, Russian Fed, Saudi Arabia, Taiwan, France, Thailand, and Italy.
- Lot, serial or model codes
- Serial numbers 250157-250190.
- 510(k) numbers
- ["K081148"]
- PMA numbers
- not on file
- Product code
- IYO
- Firm FEI
- 2936884
- Firm city
- Mountain View
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as