Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc.: ACUSON S2000 Automated Breast Volume Scanner (ABVS), model number 10434730. The affected sub-components are #10434968 - Boom Welding and 10434969 - Arm Welding. The S2000 ABVS ultrasound System for imaging the breast allows volumetric acquisition of the breast image.

Recall number
Z-1063-2015
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
ACUSON S2000 Automated Breast Volume Scanner (ABVS), model number 10434730. The affected sub-components are #10434968 - Boom Welding and 10434969 - Arm Welding. The S2000 ABVS ultrasound System for imaging the breast allows volumetric acquisition of the breast image.
Status
Terminated
Initiated
2015-01-12
Posted
2015-02-07
Terminated
2015-06-10
Reason
There is a low probability the scanner arm will become completely detached from the scanner column.
Root cause
Vendor change control
Action
Siemens sent a Customer Safety Advisory Notification to all affected sites on January 12, 2015, certified mail, with return receipt. The letter identified the affected product, the issue, and potential risk. Customers were asked to immediately discontinue use of the affected product. A Customer Service Engineer will contact customers to schedule an appointment to make the necessary repairs. Questions should be directed to a local service support person for information regarding timelines and status. Customers are to share the information with all personnel, who need to be aware of the issue.
Quantity
24 units
Distribution
Worldwide Distribution -- US, including the states of TX and MN, and the territory of P.R.; and, the countries of China, Germany, Russian Fed, Saudi Arabia, Taiwan, France, Thailand, and Italy.
Lot, serial or model codes
Serial numbers 250157-250190.
510(k) numbers
["K081148"]
PMA numbers
not on file
Product code
IYO
Firm FEI
2936884
Firm city
Mountain View
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register