FDA Medical Device Recalls (openFDA)
Covidien LP: Covidien Endo GIA AutoSuture Universal Loading Unit, 30mm-2.0mm Product Code: 030450 Applications in abdominal, gynecologic, pediatric anti thoracic surgery for resection, transection and creation of anastomosis.
- Recall number
- Z-1063-2013
- Recalling firm
- Covidien LP
- Product
- Covidien Endo GIA AutoSuture Universal Loading Unit, 30mm-2.0mm Product Code: 030450 Applications in abdominal, gynecologic, pediatric anti thoracic surgery for resection, transection and creation of anastomosis.
- Status
- Terminated
- Initiated
- 2013-02-11
- Posted
- 2013-04-05
- Terminated
- 2014-03-20
- Reason
- Single Use Loading Unit contained two staples loaded in each cartridge pocket and may result in difficulty firing and removing the device from the application site, which may require medical intervention
- Root cause
- Process control
- Action
- Covidien notified customers with an Urgent Medical Device Recall notification on February 11, 2013. The letter identified the affected product and the reason for the recall. Accounts were requested to quarantine and discontinue use of the affected devices, and notify Universal Loading Unit 30mm-2.0mm users of this recall. Customers should follow the instructions provided in order to return the affected product. Customers are to complete and return the Product Return Form regardless if they have any affected product or not. For questions regarding the RGA I return process, please contact Covidien Customer Service, M-F, 8 am - 6:30 pm ET at (800) 962-9888, option 1, and then option 2.
- Quantity
- 6,751 units
- Distribution
- Worldwide Distribution -- USA, Asia Pacific, Canada, Europe, and Japan.
- Lot, serial or model codes
- Lot Codes: N2F0563LX, N2F0341LX, N2F0725LX, and N2E0605LX.
- 510(k) numbers
- ["K111825"]
- PMA numbers
- not on file
- Product code
- GDW
- Firm FEI
- 1219930
- Firm city
- North Haven
- Firm state
- CT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28