FDA Medical Device Recalls (openFDA)
Inrad, Inc: Inrad sample master aspiration biopsy needle with hiliter sharp bevel needle, 22 ga x 2 3/8'' (6 cm), model 41222, Inrad, Inc. Kentwood, MI. The device is intended for use during percutaneous biopsy and drainage procedures to obtain tissues or fluid specimens.
- Recall number
- Z-1063-2011
- Recalling firm
- Inrad, Inc
- Product
- Inrad sample master aspiration biopsy needle with hiliter sharp bevel needle, 22 ga x 2 3/8'' (6 cm), model 41222, Inrad, Inc. Kentwood, MI. The device is intended for use during percutaneous biopsy and drainage procedures to obtain tissues or fluid specimens.
- Status
- Terminated
- Initiated
- 2010-12-20
- Posted
- 2011-02-08
- Terminated
- 2011-02-09
- Reason
- This lot of product has the wrong expiration date on the carton label. The pouches contained the correct expiration date.
- Root cause
- Employee error
- Action
- The firm, Inrad Inc., sent an "URGENT FIELD CORRECTION" notice dated December 20, 2010 to its customers. The notice described the product, problem and actions to be taken by the customers. The customers were instructed to immediately examine their inventory for the affected product. If the recalled product was in stock, the customers were told to removed the affected product from the carton and dispose of carton; to disregard carton label expiration date and reference the expiration date appearing on the pouch label only, and to complete, sign and return the attached "Carton Expiration Date" INFORMATION FORM via fax to 616-301-779 or mail to Inrad Inc., ATTN: Director RA/QA, 4375 Donker Court SE, Kentwood, MI 49512. If you have any additional questions about this action, please call 800-558-4647 extension 102.
- Quantity
- 250
- Distribution
- Nationwide Distribution: USA including states of: FL, GA, KY, NH, NJ, NM, PA , SC, WA, and WI.
- Lot, serial or model codes
- 6100-7001.
- 510(k) numbers
- ["K813368"]
- PMA numbers
- not on file
- Product code
- GAA
- Firm FEI
- 3001237677
- Firm city
- Kentwood
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28