FDA Medical Device Recalls (openFDA)
AGFA Corp.: Agfa DX-D Imaging Package - (The detector power cables described in the Report of Correction and Removal are connected to portable detector panels used in the DX-D systems) Product Usage: DX-D Imaging Package ha the same intended use as the predicative devices: namely to provide diagnostic quality images to aid the physician with diagnosis. The device uses Agfas fla panel detectors with amorphous silicon scintillators and NX workstations with MUSICA imaeg processing to create radiographic images of the skeleton (including skull, spinal column and extremities), chest, abdomen and other body parts. The product is not indicated for use in mammography.
- Recall number
- Z-1062-2012
- Recalling firm
- AGFA Corp.
- Product
- Agfa DX-D Imaging Package - (The detector power cables described in the Report of Correction and Removal are connected to portable detector panels used in the DX-D systems) Product Usage: DX-D Imaging Package ha the same intended use as the predicative devices: namely to provide diagnostic quality images to aid the physician with diagnosis. The device uses Agfas fla panel detectors with amorphous silicon scintillators and NX workstations with MUSICA imaeg processing to create radiographic images of the skeleton (including skull, spinal column and extremities), chest, abdomen and other body parts. The product is not indicated for use in mammography.
- Status
- Terminated
- Initiated
- 2012-01-16
- Posted
- 2012-02-19
- Terminated
- 2012-05-10
- Reason
- The firm received a complaint that when using the DX-D Mobile DR X-Ray unit they noticed the detector cable was hot.
- Root cause
- Device Design
- Action
- AGFA Corp. sent an "URGENT FIELD SAFETY NOTICE" letter dated January 16, 2012 to affected consignees. The letter described the affected product, concerns, issues and actions to be taken. Customers have been requested to place the Varian warning label included in a highly visible and suitable position on the cable of the portable detector. Customers were instructed to complete and return the feedback form by fax to 864-421-1664, as soon as possible. For questions about this matter contact your local AGFA HealthCare Technical Support at 877-777-2432.
- Quantity
- 42 units
- Distribution
- Nationwide Distribution (USA) including the states of: FL, IL, IN, KS, MI, MN, MS, NE,NY, OH, SC and WV.
- Lot, serial or model codes
- Model Numbers DX-D 20G and DX-D 20C; Catalog Numbers ABC Code 5PYTM and 5PYUO
- 510(k) numbers
- ["K092669"]
- PMA numbers
- not on file
- Product code
- MQB
- Firm FEI
- 3001236302
- Firm city
- Greenville
- Firm state
- SC
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28