AGFA Corp.: Agfa DX-D Imaging Package - (The detector power cables described in the Report of Correction and Removal are connected to portable detector panels used in the DX-D systems) Product Usage: DX-D Imaging Package ha the same intended use as the predicative devices: namely to provide diagnostic quality images to aid the physician with diagnosis. The device uses Agfas fla panel detectors with amorphous silicon scintillators and NX workstations with MUSICA imaeg processing to create radiographic images of the skeleton (including skull, spinal column and extremities), chest, abdomen and other body parts. The product is not indicated for use in mammography.

Recall number
Z-1062-2012
Recalling firm
AGFA Corp.
Product
Agfa DX-D Imaging Package - (The detector power cables described in the Report of Correction and Removal are connected to portable detector panels used in the DX-D systems) Product Usage: DX-D Imaging Package ha the same intended use as the predicative devices: namely to provide diagnostic quality images to aid the physician with diagnosis. The device uses Agfas fla panel detectors with amorphous silicon scintillators and NX workstations with MUSICA imaeg processing to create radiographic images of the skeleton (including skull, spinal column and extremities), chest, abdomen and other body parts. The product is not indicated for use in mammography.
Status
Terminated
Initiated
2012-01-16
Posted
2012-02-19
Terminated
2012-05-10
Reason
The firm received a complaint that when using the DX-D Mobile DR X-Ray unit they noticed the detector cable was hot.
Root cause
Device Design
Action
AGFA Corp. sent an "URGENT FIELD SAFETY NOTICE" letter dated January 16, 2012 to affected consignees. The letter described the affected product, concerns, issues and actions to be taken. Customers have been requested to place the Varian warning label included in a highly visible and suitable position on the cable of the portable detector. Customers were instructed to complete and return the feedback form by fax to 864-421-1664, as soon as possible. For questions about this matter contact your local AGFA HealthCare Technical Support at 877-777-2432.
Quantity
42 units
Distribution
Nationwide Distribution (USA) including the states of: FL, IL, IN, KS, MI, MN, MS, NE,NY, OH, SC and WV.
Lot, serial or model codes
Model Numbers DX-D 20G and DX-D 20C; Catalog Numbers ABC Code 5PYTM and 5PYUO
510(k) numbers
["K092669"]
PMA numbers
not on file
Product code
MQB
Firm FEI
3001236302
Firm city
Greenville
Firm state
SC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register