Shape Medical Systems, Inc: Shape-HF Cardiopulmonary Testing System Disposable Patient Interface (DPI). Part Number 0004-4001. The Shape-HF Cardiopulmonary Exercise Testing System is a stationary device that monitors parameters during laboratory or clinical conditions.
- Recall number
- Z-1060-2015
- Recalling firm
- Shape Medical Systems, Inc
- Product
- Shape-HF Cardiopulmonary Testing System Disposable Patient Interface (DPI). Part Number 0004-4001. The Shape-HF Cardiopulmonary Exercise Testing System is a stationary device that monitors parameters during laboratory or clinical conditions.
- Status
- Terminated
- Initiated
- 2014-12-23
- Posted
- 2015-02-05
- Terminated
- 2015-04-14
- Reason
- This recall has been initiated because use of the impacted DPIs could result in erroneous testing results potentially leading to incorrect diagnosis and incorrect treatment. The magnitude of error for VO2 or VCO2 parameters could be approximately 20% with use of the impacted product.
- Root cause
- Under Investigation by firm
- Action
- An Urgent Medical Device Recall letter was mailed to consignees on 12/23/2014. The letter identified affected product, described the issue and asked for product to be returned. Shape Technical Service will contact customers to assist in the return of product. A response form was asked to be returned. Customers with questions can call Shape Technical Service at 1-888-906-6266.
- Quantity
- 564 devices
- Distribution
- Distributed in the states of AZ, FL, GA, MA, MN, NJ. NY, PA, TX, and WI.
- Lot, serial or model codes
- Lot Numbers: 1412001, 1412002, 1412003 and 1412004
- 510(k) numbers
- ["K090722"]
- PMA numbers
- not on file
- Product code
- BTY
- Firm FEI
- 3008072932
- Firm city
- Saint Paul
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28