FDA Medical Device Recalls (openFDA)
Lucero Medical LLC: ENDURAMESH CORPECTOMY SPACER SYSTEM 013MM, 017MM, 023MM by varying heights. Dist. By: Lucero Medical, PO Box 67, Richfield, Ohio 44826 Product Usage: Device Description: The Enduramesh is a vertebral body replacement system which provides structural support of the vertebral bodies following an anterior or far lateral corpectomy, aiding in spinal fusion. The device may be used with a cleared supplemental fixation system based on surgeon discretion and preference. The Enduramesh vertebral body replacement system consists of a single cylindrically shaped titanium mesh cage. The hollow core of the cage allows for packing bone graft and the circular holes throughout the device promote bone fusions. Various diameters and heights are available to accommodate variability among patients. Indications for Use: The Enduramesh is a vertebral body replacement system intended for use in the thoracic and lumbar spine (T1-S1) to replace a collapsed, damaged or unstable vertebral body due to tumor or trauma/fracture. The device is intended for use with either allograft or autoograft. The device may be used with a cleared supplemental fixation system based on surgeon discretion and preference.
- Recall number
- Z-1059-2012
- Recalling firm
- Lucero Medical LLC
- Product
- ENDURAMESH CORPECTOMY SPACER SYSTEM 013MM, 017MM, 023MM by varying heights. Dist. By: Lucero Medical, PO Box 67, Richfield, Ohio 44826 Product Usage: Device Description: The Enduramesh is a vertebral body replacement system which provides structural support of the vertebral bodies following an anterior or far lateral corpectomy, aiding in spinal fusion. The device may be used with a cleared supplemental fixation system based on surgeon discretion and preference. The Enduramesh vertebral body replacement system consists of a single cylindrically shaped titanium mesh cage. The hollow core of the cage allows for packing bone graft and the circular holes throughout the device promote bone fusions. Various diameters and heights are available to accommodate variability among patients. Indications for Use: The Enduramesh is a vertebral body replacement system intended for use in the thoracic and lumbar spine (T1-S1) to replace a collapsed, damaged or unstable vertebral body due to tumor or trauma/fracture. The device is intended for use with either allograft or autoograft. The device may be used with a cleared supplemental fixation system based on surgeon discretion and preference.
- Status
- Terminated
- Initiated
- 2012-01-11
- Posted
- 2012-02-17
- Terminated
- 2014-04-11
- Reason
- During an FDA inspection the firm was informed that the device marketed is not the device cleared for the 510K associated with the device. The product is sold unlabeled.
- Root cause
- Device Design
- Action
- Lucero Medical LLC made a personal visit to facilities in regards to an Urgent Medical Device Recall as of January 12, 2012. The firm instructed the customer that they were removing and quarantining all product from the facility. If you have any further questions, please contact Lucero Medical, PO Box 69, Richfield, Ohio 44826, Director of Quality Assurance/Natalie Williams/ (440) 670-9583
- Quantity
- 190 units
- Distribution
- USA Nationwide Distribution - in the state of OH.
- Lot, serial or model codes
- Part # 762.010, Lot # 110738; Part # 762.019, Lot # 90611, Lot #, 100308; Part # 762.021, Lot # 100308, Lot #, 110106; Part # 762.023, Lot # 90611, Lot # 100218, Lot # 100706, Lot # 110709; Part # 762.025, Lot # 90611, Lot # 100218, Lot #, 110106; Part # 762.027, Lot # 100706; Part # 762.029, Lot # 90611; Part # 762.033, Lot # 90611, Lot # 91002; Part # 762.041, Lot # 90611; Part # 762.043, Lot # 90611; Part # 762.045, Lot # 90611; Part # 762.047, Lot # 90611; Part # 762.049, Lot # 90611; Part # 762.051, Lot # 90611; Part # 762.053, Lot # 90611; Part # 762.090-2.0, Lot # 100206; Part # 762.090-2.5, Lot # 100206; Part # 762.237, Lot # 91002; Part # 762.243, Lot # 91018; Part # 762.270, Lot # 91018; Part # 762.290, Lot # 91018; Part # 762.731, Lot # 91018; Part # 762.735, Lot # 91002; Part # 762.767, Lot # 91018 & Part # 762.790, Lot # 91018.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- MQP
- Firm FEI
- 3008775608
- Firm city
- Hinckley
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28