Covidien LP (formerly Nellcor Puritan Bennett Inc.): Puritan Bennett 980 Ventilator System, Model No. PB980 Ventilator (980xxxxxxxx), The Puritan BennettTM 980 Ventilator System is designed for use on Neonatal (NICU) through Adult patient populations who require respiratory support or mechanical ventilation and weigh a minimum of 0.3 kg (0.66 lb). It is suitable for service in a hospital (institutions) and intra-hospital transport to provide continuous positive pressure ventilator support, delivered invasively or noninvasively, to patients who require the following types of ventilator support: "Positive Pressure Ventilation, delivered invasively (via endotracheal tube or trach tube) or non-invasively (via mask or nasal prongs) "Assist/ Control, SIMV or Spontaneous modes of ventilation.

Recall number
Z-1058-2015
Recalling firm
Covidien LP (formerly Nellcor Puritan Bennett Inc.)
Product
Puritan Bennett 980 Ventilator System, Model No. PB980 Ventilator (980xxxxxxxx), The Puritan BennettTM 980 Ventilator System is designed for use on Neonatal (NICU) through Adult patient populations who require respiratory support or mechanical ventilation and weigh a minimum of 0.3 kg (0.66 lb). It is suitable for service in a hospital (institutions) and intra-hospital transport to provide continuous positive pressure ventilator support, delivered invasively or noninvasively, to patients who require the following types of ventilator support: "Positive Pressure Ventilation, delivered invasively (via endotracheal tube or trach tube) or non-invasively (via mask or nasal prongs) "Assist/ Control, SIMV or Spontaneous modes of ventilation.
Status
Terminated
Initiated
2015-01-12
Posted
2015-02-05
Terminated
2016-01-06
Reason
Covidien is issuing a voluntary field action for all Puritan Bennett 980 ventilators due to occasional GUI transient resets that last approximately 30 seconds.
Root cause
Software design
Action
Covidien sent an Urgent Field Safety Notice on January 12, 2015, to all affected customers. Customers were requested to forward the recall letter to anyone they may have transferred or assigned for use one of the potentially affected ventilators. Covidien is updating software that will modify ventilator alarms and status display messages during a GUI reset. The Puritan Bennett 980 Ventilator Operator Manual will be updated to delete,replace ventilator during a GUI transient reset, and to inform users to monitor the secondary status display until the primary display refreshes. Covidien will communicate with customers when the software and operator manual updates are available. Customers with questions were instructed to contact Technical Support Department at 800-255-6774, option 4, and then option 1. For questions regarding this recall call 303-876-8870.
Quantity
418 units
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Canada.
Lot, serial or model codes
All Puritan Bennett 980 ventilators
510(k) numbers
["K131252"]
PMA numbers
not on file
Product code
CBK
Firm FEI
2936999
Firm city
Boulder
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register