FDA Medical Device Recalls (openFDA)
Stryker Howmedica Osteonics Corp.: Omnifit M/S Dual Geom Shell 58 MM; Stryker Orthopaedics, Uses in primary and revision total hip arthroplasty which are indicated for painful, disabling joint disease resulting from degenerative arthritis, rheumatoid arthritis, post traumatic arthritis or late stage avascular necrosis.
- Recall number
- Z-1054-2011
- Recalling firm
- Stryker Howmedica Osteonics Corp.
- Product
- Omnifit M/S Dual Geom Shell 58 MM; Stryker Orthopaedics, Uses in primary and revision total hip arthroplasty which are indicated for painful, disabling joint disease resulting from degenerative arthritis, rheumatoid arthritis, post traumatic arthritis or late stage avascular necrosis.
- Status
- Terminated
- Initiated
- 2010-08-02
- Posted
- 2011-02-07
- Terminated
- 2012-11-09
- Reason
- Stryker Orthopaedics determined that specific lots of the Omnifit Acetabular Shells may have a screw hole wall thickness that falls above the design specification.
- Root cause
- Device Design
- Action
- The firm, Stryker Orthopaedics, sent an "URGENT PRODUCT RECALL" letter dated August 2, 2010, via Federal Express to all customers. The letter described the product, problem and action to be taken by the customers. The customers were instructed to examine their inventory to identify the affected product; retrieve all affected product and return it their branch or agency warehouse for reconciliation; reconcile all product(s) on the attached Product Recall Acknowledgement Form and fax to (201) 831-6069, and return the affected product via Inter-Org Transfer and ship to the attention of Regulatory Compliance, Stryker Orthopaedics, 325 Corporate Drive, Mahwah, NJ 07430. If you have any questions, feel free to contact the Manager, Divisional Regulatory Compliance at (201) 972-2100.
- Quantity
- not on file
- Distribution
- Worldwide distribution: USA including states of: AL, AR, CA, FL, IN, ME, MI, MS, NC, NJ, OK, OR, WA , and WI; and countries including: China, Hong Kong and Korea.
- Lot, serial or model codes
- Catalog number 2035-0058, Lot # SS3034A.
- 510(k) numbers
- ["K912654"]
- PMA numbers
- not on file
- Product code
- MEH
- Firm FEI
- 3003070421
- Firm city
- Mahwah
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28