FDA Medical Device Recalls (openFDA)
Baxter Healthcare Corp.: Clintec Micromix Compounder, an I.V. solution compounder; Baxter Healthcare Corporation, Clintec Nutrition Division, Deerfield, IL 60015 USA; product code 2M8290 - 1.0 mL accuracy, and product code 2M8290L - less than 1.0 mL accuracy
- Recall number
- Z-1054-04
- Recalling firm
- Baxter Healthcare Corp.
- Product
- Clintec Micromix Compounder, an I.V. solution compounder; Baxter Healthcare Corporation, Clintec Nutrition Division, Deerfield, IL 60015 USA; product code 2M8290 - 1.0 mL accuracy, and product code 2M8290L - less than 1.0 mL accuracy
- Status
- Terminated
- Initiated
- 2004-04-26
- Posted
- 2004-07-20
- Terminated
- 2006-04-11
- Reason
- The Micromix compounder may have ingredient delivery outside of the stated accuracy limits if the Vlier pin is not properly adjusted
- Root cause
- Other
- Action
- Baxter began telephoning their customers on 4/26/04, advising them that an improperly adjusted Vlier pin could result in an ingredient delivery outside of stated accuracy limits. The accounts were requested to immediately stop using the Micromix Compounder and return the device to Baxter, and to evaluate all TPN formualtions that have been compounded but not administered to determine if any ingredients were delivered at a programmed volume of less than 1 mL. The account was advised to discard those solutions and recompound them using manual techniques. Follow-up recall letters were faxed to the accounts on the same date, reiterating the telephone calls. Baxter sent second recall letters dated 2/16/05 to the 15 customers who had not yet returned their Micromix Compounder as of 2/16/05, along with a copy of the 4/26/04 recall letter. The customers were informed of the new issue with the Micromix Compounder not recognizing decimal points for delivery volumes below 1 mL, and were again requested to return the Micromix Compounder and the handheld keypad to Baxter so that they may be corrected when the Micromix is reintroduced the market.
- Quantity
- 500 devices
- Distribution
- Nationwide and internationally through Baxter subsidiaries to the United Kingdom, Germany, Chile, Canada, Brazil, El Salvador, Taiwan and Korea.
- Lot, serial or model codes
- product codes 2M8290 and 2M8290L, all serial numbers The recall extension concerns those units that had not been returned as of 2/16/05: serial numbes 10025, 11163, 11263, 11298, 11254, 11280, 10075, 11279, 11265, 11264, 10065, 11251, 11065, T12054, 10051
- 510(k) numbers
- ["K955622"]
- PMA numbers
- not on file
- Product code
- NEP--
- Firm FEI
- 1416980
- Firm city
- Round Lake
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28