Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corp.: Clintec Micromix Compounder, an I.V. solution compounder; Baxter Healthcare Corporation, Clintec Nutrition Division, Deerfield, IL 60015 USA; product code 2M8290 - 1.0 mL accuracy, and product code 2M8290L - less than 1.0 mL accuracy

Recall number
Z-1054-04
Recalling firm
Baxter Healthcare Corp.
Product
Clintec Micromix Compounder, an I.V. solution compounder; Baxter Healthcare Corporation, Clintec Nutrition Division, Deerfield, IL 60015 USA; product code 2M8290 - 1.0 mL accuracy, and product code 2M8290L - less than 1.0 mL accuracy
Status
Terminated
Initiated
2004-04-26
Posted
2004-07-20
Terminated
2006-04-11
Reason
The Micromix compounder may have ingredient delivery outside of the stated accuracy limits if the Vlier pin is not properly adjusted
Root cause
Other
Action
Baxter began telephoning their customers on 4/26/04, advising them that an improperly adjusted Vlier pin could result in an ingredient delivery outside of stated accuracy limits. The accounts were requested to immediately stop using the Micromix Compounder and return the device to Baxter, and to evaluate all TPN formualtions that have been compounded but not administered to determine if any ingredients were delivered at a programmed volume of less than 1 mL. The account was advised to discard those solutions and recompound them using manual techniques. Follow-up recall letters were faxed to the accounts on the same date, reiterating the telephone calls. Baxter sent second recall letters dated 2/16/05 to the 15 customers who had not yet returned their Micromix Compounder as of 2/16/05, along with a copy of the 4/26/04 recall letter. The customers were informed of the new issue with the Micromix Compounder not recognizing decimal points for delivery volumes below 1 mL, and were again requested to return the Micromix Compounder and the handheld keypad to Baxter so that they may be corrected when the Micromix is reintroduced the market.
Quantity
500 devices
Distribution
Nationwide and internationally through Baxter subsidiaries to the United Kingdom, Germany, Chile, Canada, Brazil, El Salvador, Taiwan and Korea.
Lot, serial or model codes
product codes 2M8290 and 2M8290L, all serial numbers The recall extension concerns those units that had not been returned as of 2/16/05: serial numbes 10025, 11163, 11263, 11298, 11254, 11280, 10075, 11279, 11265, 11264, 10065, 11251, 11065, T12054, 10051
510(k) numbers
["K955622"]
PMA numbers
not on file
Product code
NEP--
Firm FEI
1416980
Firm city
Round Lake
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28