FDA Medical Device Recalls (openFDA)
Philips North America, LLC: Urodiagnost, Product No. 708033 Product Usage: The Philips UroDiagnost MRF can be used to perform all urological and gynecological routine examinations including: Urograms with injection or infusion Retrograde pyelography Cystography and urethrography Micturition cystography Cystoscopy, endopscopy Transurethral resection Percutaneous nephroscopy Urethrorenoscopy Percutaneous nephrolithotomy Serial angiography with image intensifier and DSI Tomography
- Recall number
- Z-1053-2019
- Recalling firm
- Philips North America, LLC
- Product
- Urodiagnost, Product No. 708033 Product Usage: The Philips UroDiagnost MRF can be used to perform all urological and gynecological routine examinations including: Urograms with injection or infusion Retrograde pyelography Cystography and urethrography Micturition cystography Cystoscopy, endopscopy Transurethral resection Percutaneous nephroscopy Urethrorenoscopy Percutaneous nephrolithotomy Serial angiography with image intensifier and DSI Tomography
- Status
- Terminated
- Initiated
- 2019-02-28
- Posted
- not on file
- Terminated
- 2020-08-03
- Reason
- The locking plate may be missing from the system, which may result in unexpected stand/table tilt movement that cannot be stopped by the user.
- Root cause
- Component design/selection
- Action
- On February 28, 2019, the firm notified customers via an Urgent Field Safety Notice letter. The letter informed customers that the affected systems would be inspected for missing locking plates. Customers were asked to place a copy of the Field Safety Notice with the documentation of the system until the system has been inspected by Philips. Philips will inspect all potentially affected systems free of charge. If a locking plate is found missing, Philips will secure the system until a locking plate is installed so that use of the system can continue. You will be contacted by your local Philips representative to schedule the inspection. Philip's action will start effective from March 2019. If you need further information or support concerning this issue, please contact your local Philips representative.
- Quantity
- 1969
- Distribution
- US Nationwide domestic distribution, and worldwide international distribution.
- Lot, serial or model codes
- Serial codes 1 2 3 4 5 7 8 11 13 14 15 16 17 18 21 48 59 60 64
- 510(k) numbers
- ["K911098"]
- PMA numbers
- not on file
- Product code
- IXR
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28