FDA Medical Device Recalls (openFDA)

Philips North America, LLC: Urodiagnost, Product No. 708033 Product Usage: The Philips UroDiagnost MRF can be used to perform all urological and gynecological routine examinations including: Urograms with injection or infusion Retrograde pyelography Cystography and urethrography Micturition cystography Cystoscopy, endopscopy Transurethral resection Percutaneous nephroscopy Urethrorenoscopy Percutaneous nephrolithotomy Serial angiography with image intensifier and DSI Tomography

Recall number
Z-1053-2019
Recalling firm
Philips North America, LLC
Product
Urodiagnost, Product No. 708033 Product Usage: The Philips UroDiagnost MRF can be used to perform all urological and gynecological routine examinations including: Urograms with injection or infusion Retrograde pyelography Cystography and urethrography Micturition cystography Cystoscopy, endopscopy Transurethral resection Percutaneous nephroscopy Urethrorenoscopy Percutaneous nephrolithotomy Serial angiography with image intensifier and DSI Tomography
Status
Terminated
Initiated
2019-02-28
Posted
not on file
Terminated
2020-08-03
Reason
The locking plate may be missing from the system, which may result in unexpected stand/table tilt movement that cannot be stopped by the user.
Root cause
Component design/selection
Action
On February 28, 2019, the firm notified customers via an Urgent Field Safety Notice letter. The letter informed customers that the affected systems would be inspected for missing locking plates. Customers were asked to place a copy of the Field Safety Notice with the documentation of the system until the system has been inspected by Philips. Philips will inspect all potentially affected systems free of charge. If a locking plate is found missing, Philips will secure the system until a locking plate is installed so that use of the system can continue. You will be contacted by your local Philips representative to schedule the inspection. Philip's action will start effective from March 2019. If you need further information or support concerning this issue, please contact your local Philips representative.
Quantity
1969
Distribution
US Nationwide domestic distribution, and worldwide international distribution.
Lot, serial or model codes
Serial codes 1 2 3 4 5 7 8 11 13 14 15 16 17 18 21 48 59 60 64
510(k) numbers
["K911098"]
PMA numbers
not on file
Product code
IXR
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register