FDA Medical Device Recalls (openFDA)
Cooper Surgical, Inc.: LEEP REDIKIT Product No. 6060 In a Loop Electrosurgical Excision Procedure, a thin, low-voltage electrified wire loop is used to cut out abnormal cervical tissue. The cartridges of Polocaine and Xylocaine with Epinephrine contained in the LEEP RediKit are local anaesthetic which the physician may choose from during the procedure.
- Recall number
- Z-1053-2013
- Recalling firm
- Cooper Surgical, Inc.
- Product
- LEEP REDIKIT Product No. 6060 In a Loop Electrosurgical Excision Procedure, a thin, low-voltage electrified wire loop is used to cut out abnormal cervical tissue. The cartridges of Polocaine and Xylocaine with Epinephrine contained in the LEEP RediKit are local anaesthetic which the physician may choose from during the procedure.
- Status
- Terminated
- Initiated
- 2013-02-26
- Posted
- 2013-04-03
- Terminated
- 2014-06-12
- Reason
- LEEP RediKit Products (for use in Loop Electrosurgical Excision Procedures) have Xylocaine (Lidocaine) HCL (2%) listed on the tray lid label but this is not included in the contents of the kit, which contained Polocaine (Mepivacaine) HCL (3%) instead (which was not listed on the tray label).
- Root cause
- Labeling mix-ups
- Action
- CooperSurgical sent a Recall Notification dated February 26, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. CSI intends to track and quarantine returned product in a segregated location at its manufacturing facility to ensure its separation from acceptable product and prevent its misuse. CSI intends to render used product useless by physical destruction and to rework unused product by correcting the labeling for redistribution. If you have any one of these affected lot numbers in your current inventory, please contact our Customer Service Department at 800-243-2974, or simply complete the attached form and fax to 800-262-0105. Arrangements will be made to replace product in your possession at CooperSurgical's expense. Further instruction and disposition of product will be communicated to you at that time. CooperSurgical deeply regrets the inconvenience that this problem has caused.
- Quantity
- 4395 kits
- Distribution
- Nationwide Distribution including Puerto Rico
- Lot, serial or model codes
- Lot 120716, exp 8-31-13; Lot 116657 exp. 6-30-13; Lot 116658 exp. 5-31-13; Lot 116659 exp. 7-31-13
- 510(k) numbers
- ["K910253"]
- PMA numbers
- not on file
- Product code
- MQB
- Firm FEI
- 1216677
- Firm city
- Trumbull
- Firm state
- CT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28