Skip to the content

FDA Medical Device Recalls (openFDA)

NuVasive Inc: NuVasive MD Pulse III Multimodality System REF 1881500 Pulse Software version 369-BL.US and 369-BLN.US

Recall number
Z-1052-2023
Recalling firm
NuVasive Inc
Product
NuVasive MD Pulse III Multimodality System REF 1881500 Pulse Software version 369-BL.US and 369-BLN.US
Status
Open, Classified
Initiated
2022-07-19
Posted
2023-02-03
Terminated
not on file
Reason
Due to an incorrect security key, customers were unable to connect to remote monitoring during spinal surgery, neck dissections, thoracic surgeries, and upper and lower extremity procedures.
Root cause
Software design
Action
Between 07/19/2022 to 12/02/2022, affected customers were contacted and verbally informed during a software update that there was an issue related to the inability to remotely connect their system and does not allow physicians who is not physically in the operating room to be able to review live neuromonitoring data remotely and provide analysis to the surgeon in the operating room. The firm is planning on sending an "URGENT: MEDICAL DEVICE RECALL" Letter on 01/13/2023, to officially inform customers that a recall is being conducted due to a remote monitoring issue that may potentially not allow a physician who is not physically in the operating room to be able to review live neuromonitoring data remotely and provide analysis to the surgeon in the operating room. Customer Instructions: -Nuvasive, Inc. will contact a Representative at the hospitals to schedule the software key upgrade. -If the security key has been updated, no further action is necessary. For questions or assistance, contact Sr. Director, Global Quality at 858-336-3421 or email pyrigoyen@nuvasive.com
Quantity
7
Distribution
U.S. Nationwide distribution in the states of CA, FL, NC, OH, and TX.
Lot, serial or model codes
Part Number: 1881500 UDI Code: 0887517PLS137YJ Serial Numbers: PM00017 SM90018 SM90013 SM90021 PM00007 SM90032 SM90030
510(k) numbers
["K210574"]
PMA numbers
not on file
Product code
OLO
Firm FEI
3002980729
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register