Skip to the content

FDA Medical Device Recalls (openFDA)

Instrumentation Laboratory: HemosIL ReadiPlasTin, Part No. 0020301400

Recall number
Z-1052-2022
Recalling firm
Instrumentation Laboratory
Product
HemosIL ReadiPlasTin, Part No. 0020301400
Status
Open, Classified
Initiated
2022-04-08
Posted
not on file
Terminated
not on file
Reason
Variable and out of specification QC results with HemosIL ReadiPlasTin. IL will remove all lots from the market and convert customers to an alternative prothrombin time (PT) reagent, specifically HemosIL RecombiPlasTin 2G.
Root cause
Under Investigation by firm
Action
On April 8, 2022, the firm notified customers via email and certified mail. Customers were informed that the firm will stop shipment of product on May 9, 2022. After receipt of the material, customers must complete the conversion to HemosIL RecombiPlasTin 2G, with support as needed from your local Werfen representative, within 90 days. Until a lot of HemosIL RecombiPlasTin 2G has been received and ready for use, customers should run quality control for HemosIL ReadiPlasTin with each new reagent vial (every lot) and at least every 8 hours to identify vials with performance issues. Any product with failed quality control or quality control results which trend high should be discarded. Once customers have completed the comversion to HemosIL RecombiPlasTin 2G, any remaining unused HemosIL ReadiPlasTin should be destroyed.
Quantity
1,563 (US); 29,046 (OUS)
Distribution
US Nationwide distribution including in the states of AZ, CA, IA, IL, IN, KS, MD, ME, MI, MN, MO, NC, NE, NY, OH, OK, SC, TX, VA, WV.
Lot, serial or model codes
UDI 08426950632887 All in-date lots are affected by this recall.
510(k) numbers
["K122584"]
PMA numbers
not on file
Product code
GJS
Firm FEI
1217183
Firm city
Bedford
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register