Skip to the content

FDA Medical Device Recalls (openFDA)

Cardinal Health 200, LLC: Dover" 100% SILICONE FOLEY TRAY, LUER LOCK, 16 FR (5.3 MM), 5 CC, 2000 ML DRAIN BAG, DRAIN SPOUT; Dover" 100% SILICONE 400 ML U/M FOLEY TRAY, LUER LOCK, 16 FR (5.3 MM), 5 CC, 2000 ML DRAIN BAG; Dover" 100% SILICONE 400 ML U/M FOLEY TRAY, COUD TIP, 16 FR (5.3 MM), 5 CC, 2000 ML DRAIN BAG, CSD; Dover" 100% SILICONE SILVER-HYDROGEL COATED 400 ML U/M FOLEY TRAY, LL, 16 FR (5.3 MM), 5 CC, 2-WAY; Dover" 100% SILICONE SILVER-HYDROGEL COATED FOLEY INSERTION TRAY, 5 CC, PREP TRAY, 16 FR (5.3 MM); Dover" 100% SILICONE SILVER-HYDROGEL COATED FOLEY TRAY, 16 FR (5.3 MM), 5CC, 2-WAY, 2000ML DRAIN BAG; Dover" 100% SILICONE SILVER-HYDROGEL COATED FOLEY TRAY, 18 FR (6.0 MM), 5CC, 2-WAY, 2000ML DRAIN BAG; Dover" HYDROGEL COATED COUD TIP LATEX FOLEY TRAY, 16 FR (5.3 MM), 5 CC, 2000 ML DRAINAGE BAG; Dover" HYDROGEL COATED LATEX FOLEY TRAY, LUER-LOCK, 18 FR (6.0 MM), 5 CC, 2000 ML BAG, DRAIN SPOUT; Dover" 100% SILICONE 400 ML U/M FOLEY TRAY, LUER LOCK, 18 FR (6.0 MM), 5 CC, 2000 ML DRAIN BAG; Dover" HYDROGEL COATED COUD TIP LATEX 400 ML U/M FOLEY TRAY, 16 FR (5.3 MM), 5 CC, 2000 ML BAG

Recall number
Z-1050-2024
Recalling firm
Cardinal Health 200, LLC
Product
Dover" 100% SILICONE FOLEY TRAY, LUER LOCK, 16 FR (5.3 MM), 5 CC, 2000 ML DRAIN BAG, DRAIN SPOUT; Dover" 100% SILICONE 400 ML U/M FOLEY TRAY, LUER LOCK, 16 FR (5.3 MM), 5 CC, 2000 ML DRAIN BAG; Dover" 100% SILICONE 400 ML U/M FOLEY TRAY, COUD TIP, 16 FR (5.3 MM), 5 CC, 2000 ML DRAIN BAG, CSD; Dover" 100% SILICONE SILVER-HYDROGEL COATED 400 ML U/M FOLEY TRAY, LL, 16 FR (5.3 MM), 5 CC, 2-WAY; Dover" 100% SILICONE SILVER-HYDROGEL COATED FOLEY INSERTION TRAY, 5 CC, PREP TRAY, 16 FR (5.3 MM); Dover" 100% SILICONE SILVER-HYDROGEL COATED FOLEY TRAY, 16 FR (5.3 MM), 5CC, 2-WAY, 2000ML DRAIN BAG; Dover" 100% SILICONE SILVER-HYDROGEL COATED FOLEY TRAY, 18 FR (6.0 MM), 5CC, 2-WAY, 2000ML DRAIN BAG; Dover" HYDROGEL COATED COUD TIP LATEX FOLEY TRAY, 16 FR (5.3 MM), 5 CC, 2000 ML DRAINAGE BAG; Dover" HYDROGEL COATED LATEX FOLEY TRAY, LUER-LOCK, 18 FR (6.0 MM), 5 CC, 2000 ML BAG, DRAIN SPOUT; Dover" 100% SILICONE 400 ML U/M FOLEY TRAY, LUER LOCK, 18 FR (6.0 MM), 5 CC, 2000 ML DRAIN BAG; Dover" HYDROGEL COATED COUD TIP LATEX 400 ML U/M FOLEY TRAY, 16 FR (5.3 MM), 5 CC, 2000 ML BAG
Status
Open, Classified
Initiated
2023-12-04
Posted
2024-02-16
Terminated
not on file
Reason
Convenience kits were manufactured and distributed with components which were subsequently recalled by Nurse Assist.
Root cause
Nonconforming Material/Component
Action
Cardinal Health Issued an URGENT KIT/TRAY COMPONENT RECALL NOTICE ON 12/04/2023 via FedEx next day delivery. The notice explained the issue, potential risk to health, and requested the following actions be taken: If you locate affected product within your facility, please do the following: 1. QUARANTINE affected kits/trays. 2. ALERT your clinicians of this action. 3. AFFIX the enclosed WARNING LABEL to the front of each kit/tray so that it is clearly visible to clinicians, instructing them to remove and discard the recalled component(s). 4. NOTE if you have transferred the affected kits/trays to other departments or facilities within your system, please provide a copy of this notice and recall acknowledgement form to them. 5. DISTRIBUTOR if you have further distributed affected kits/trays, please notify your customers of this recall. 6. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of over labeled product via email GMB-FieldCorrectiveAction@cardinalhealth.com or fax to 614.652.9648. Additional labels will be sent to your facility upon request. For field action concerns, please contact the field action team at: GMB-FieldCorrectiveAction@cardinalhealth.com or call 800-292-9332. On 12/21/2023, Cardinal Health issued an URGENT: EXPANSION KIT/TRAY COMPONENT RECALL notice stating the following: As an update to Cardinal Healths initial communication on December 05, 2023, additional kit/trays are being included as part of the component recall. If you did not receive the initial communication on December 05, 2023, you were not impacted by the initial scope of the event. Instructions below are unchanged. Please reference Attachment 1 - Updated Impacted Products V2 for additional products impacted. The notice explained the issue, potential risk to health, and requested the same actions be taken as listed above in the 12/04/2023 letter.
Quantity
4620 units
Distribution
US and EMEA, Japan, Latin America
Lot, serial or model codes
a) 6146LL, UDI/DI 30884521015504 (cs), 10884521015500 (ea), Lot Numbers: 2222830564, 2234606164, 2300129064, 2300129264, 2206723564, 2206723664, 2213047764, 2213741964, 2214452664, 2216503164, 2211628464, 2225100964, 2305403264, 2207404564, 2203933464, 2209525664, 2210210364, 2210210464, 2212323064, 2215116164 b) P4P16CSD, UDI/DI 20884521143347 (cs), 10884521143340 (ea), Lot Numbers: 2231220064, 2303011264, 2309041164, 2312914064, 2319901964, 2310300564, 2322102864, 2207803464, 2212321564 c) 7006ICLL, Lot Numbers: 2215120464, 2223120264 d) 7006LL, Lot Numbers: 2207805664, 2208817564, 2212323864, 2215826764, 2223542964, 2223543064, 2231912364, 2232102364, 2206721364, 2206721464, 2210904064, 2209525964, 2213048664, 2232601264
510(k) numbers
["K024010","K780062","K990500"]
PMA numbers
not on file
Product code
EZL
Firm FEI
3001236905
Firm city
Waukegan
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register