Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc.: Siemens Dimension(R) EXL(TM) 200 (Siemens Material Number 10636929). Intended to measure a variety of analytes in human body fluids.

Recall number
Z-1048-2013
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
Siemens Dimension(R) EXL(TM) 200 (Siemens Material Number 10636929). Intended to measure a variety of analytes in human body fluids.
Status
Terminated
Initiated
2013-02-27
Posted
2013-04-01
Terminated
2014-01-14
Reason
Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined that many of these errors are due to the current limit set for System Check LOCI results. The errors have no impact on results, but they may cause customer inconvenience. Siemens h
Root cause
Under Investigation by firm
Action
Siemens initiated their recall of this product on February 27, 2013 by sending an Important Customer Notification letter to consignees by FedEx. The letter identified the product, problem, and actions to be taken by the customers.
Quantity
2207 total
Distribution
Worldwide Distribution.
Lot, serial or model codes
Siemens Material Number 10636929
510(k) numbers
["K073604"]
PMA numbers
not on file
Product code
JJE
Firm FEI
2517506
Firm city
Newark
Firm state
DE
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register